To carry out the dirty window zero offset calibration make sure the windows are
very clean. Temperature should ideally be in the range 15 – 30°C.
CS125 CALIBRATION - MENU 3
ID 0
S/N 1006
(1) Perform calibration
(2) Restore the factory calibration
(3) Perform dirty windows zero offset calibration
(4) Restore dirty windows factory calibration
(9) Refresh
(0) Return to main menu
Typing ‘3’ returns text similar to the following: -
Current values EO=3200 DO=4649 DD=995
Cal DW offset? Y/N?
Then type ‘Y’. The CS125 responds in a similar way to the following: -
Calibrating dirty window system...Please wait
DD=990 DO=4535
DD=1000 DO=4531
DD=1010 DO=4373
DD=1020 DO=4206
DD=1030 DO=3886
DD=1110 DO=2675
DD=1120 DO=2682
DD=1130 DO=2530
DD=1140 DO=2392
EO=3230 ES=371 DO=2251 DS=234 DD=1140
Press any key to exit (Not return)
Press any key and the dirty window zero offset calibration is complete.
17. Performing a firmware update
Firmware updates for the CS125 present weather sensor are performed using
Campbell Scientifics Device Configuration Utility (DevConfig) software. Please
refer to the help built into the DevConfig software for full instructions on how to
update the CS125 firmware. The pictures below show the procedure using
DevConfig.