![Campbell CS125 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/campbell/cs125/cs125_instruction-manual_3803392052.webp)
CS125 Present Weather Sensor
42
point you will not be able to exit until the test is complete. However, power
cycling the unit at this point will have no adverse effect on the sensor.
Once you have started the tests you will be asked for the CS125 calibrator serial
number and coefficient with a confirmation at each step giving you the chance to
correct typing mistakes.
When asked for confirmation you do not need to press return after
you type ‘y’.
Figure 16-1. Calibration disk
NOTE
Starting calibration.
Input the CS125 calibrator serial number ->E2002
Is E2002 correct? (Y/N)?
Input the CS125 calibrator constant ->28.8
Is 28.8 correct? (Y/N)?
Place one calibration bung into each hood, then
press any key.
Do you want to perform a calibration Y/N?