User Manual
23
Dim Rs, Vs_Vx
'CNR4 sensitivities: refer to the Certificate of Calibration from Kipp & Zonen for sensitivity values
'for each probes, and enter them below.
Const pyranometer_up_sensitivity = 15.35
'unique sensitivity for upper pyranometer
'(microV/W/m^2)
Const pyranometer_dn_sensitivity = 15.41
'unique sensitivity for lower pyranometer
'(microV/W/m^2)
Const pyrgeometer_up_sensitivity = 8.50
'unique sensitivity for upper pyrgeometer
'(microV/W/m^2)
Const pyrgeometer_dn_sensitivity = 7.09
'unique sensitivity for lower pyrgeometer
'(microV/W/m^2)
'CNR4 multipliers
Public cnr4_mult(4)
Const pyranometer_up_mult = 1000/pyranometer_up_sensitivity
'(W/m^2/mV)
Const pyranometer_dn_mult = 1000/pyranometer_dn_sensitivity
'(W/m^2/mV)
Const pyrgeometer_up_mult = 1000/pyrgeometer_up_sensitivity
'(W/m^2/mV)
Const pyrgeometer_dn_mult = 1000/pyrgeometer_dn_sensitivity
'(W/m^2/mV)
DataTable (cnr4_data,True,-1)
DataInterval
(0,60,Min,10)
CardOut (1,-1)
Minimum
(1,batt_volt,FP2,0,False)
Sample
(1,logger_temp,FP2)
Average
(4,cnr4(1),IEEE4,False)
Average
(1,cnr4_T_C,IEEE4,False)
Average
(1,cnr4_T_K,IEEE4,False)
Average
(1,long_up_corr,IEEE4,False)
Average
(1,long_dn_corr,IEEE4,False)
Average
(1,Rs_net,IEEE4,False)
Average
(1,Rl_net,IEEE4,False)
Average
(1,albedo,IEEE4,False)
Average
(1,Rn,IEEE4,False)
EndTable
DataTable (cnr4_ts,True,-1)
DataInterval
(0,1,Sec,10)
CardOut (1,-1)
Sample
(4,cnr4(1),IEEE4)
Sample
(1,cnr4_T_K,IEEE4)
EndTable
BeginProg
'Load the multiplier values for the CNR4
cnr4_mult(1) = pyranometer_up_mult
cnr4_mult(2) = pyranometer_dn_mult
cnr4_mult(3) = pyrgeometer_up_mult
cnr4_mult(4) = pyrgeometer_dn_mult