05103, 05103-45, 05106, 015108, 05108-45 and 05305 R.M. Young Wind Monitors
B-2
B.2.1 Removing the 1M
Resistor from the Wind Monitor Circuit
The 1M
resistor (labelled R3) is included within the Wind Monitor circuit for
use with Campbell Scientific dataloggers. When the Wind Monitor is directly
connected to the Wind Tracker, this resistor is not required and must be removed
(or disabled).
Identify the correct resistor by referring to Figure B-1, below. It is recommended
that the resistor wire is snipped at the top edge and bent away from the resistor
itself, as shown in Figure B-2, effectively removing the resistor from the circuit. In
this way it is a fairly simple job to reconnect the resistor, should this be required at
a future date (so that you could use the Wind Monitor directly connected to a
datalogger, for example), by repositioning the bent wire and soldering the
connection.
Figure B-1 Location of 1M
Resistor on Circuit Board
If you have any problems concerning the location or removal of the
1M
resistor, please contact Campbell Scientific for advice.
NOTE
Cut the 1M
resistor wire in this
vicinity and bend away.
Circuit Board
Figure B-2 Disconnecting the Resistor from the Circuit
Location of 1M
resistor –
cut wire and bend to
disconnect resistor from circuit
as shown in Figure B-2