GW3060/4090 – User Manual
41
9.7 DRYING AIR FILTER(GW4090)
The air filter of the drying system must be periodically checked and changed. It is located in the
upper part of the detergent pump compartment.
The mat filter installed possesses 98% DOP efficiency and should be replaced once a year.
This operation is simple to do and the user can carry it out after ordering the spare filter.
Unscrew the 8 screws that hold the filter in place in order to disassemble the filter holder and install
the new filter.
9.8 EXTERNAL CLEANING
The external parts of the appliance must be periodically cleaned.
The front panel can be cleaned with a soft cloth, using water and a delicate detergent (such as the
ones used for crockery) in diluted form.
Do not use alcohol, solvents or the ammonia-based products used for cleaning windows.
The steel surfaces can be cleaned with a stainless steel cleanser.
If there are heavy incrustations, use a descaler product to remove water hardness markings.
In places where strong acids are used in the laboratory (hydrochloric/sulphuric/nitric), it is
advisable to clean the machine with a soft cloth soaked in paraffin oil. This film will protect the steel
from acid vapour aggression.
9.9 WHEN TO SERVICE THE APPLIANCE
The user should call Camlab Service Dept every 12 months for preventive maintenance purposes
and for the visual inspection required for safety reasons.
Make sure that there is a clear area of at least 1m
2
in front of the machine before
proceeding with any maintenance/repair work.
10 INSTALLED MACHINE SET-UP
Model machine
: ________________________________________________________
Machine serial number
: __________________________________________________