GW3060/4090 – User Manual
36
open/closed, water pressure,
connection pipe, etc.)
24
NO WATER IN CHAMBER
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.) Is the drain
pipe too low?
25
NO PRESS IN HYD SYS: FOAM
Is wash pump running?
Check for foam in chamber.
If foam is present, carry out
RESET, add anti-foam soln and
run Prog 1.
29
WASHING CHAMBER DRAINAGE
FAILURE
Possible drain pump failure or
blockage.Make sure that the drain
pipe is not kinked.
30
SAFETY CHAMBER LEVEL EXCEEDED
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
31
SAFETY LEVEL FAILED
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
32 SUMP
FULL
Make a Reset cycle. Repeat the
cycle. Call the Technical
Assistance Service if the fault
persists.
33
LACK OF WATER IN THE STEAM
CONDENSER
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
34
CONDENSER DRAINAGE FAILED
Make sure that the drain pipe is
positioned as indicated in the
manual.
35
CONDENSER LEVEL SWITCH FAILED
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
36
CONDENSER DRAINING PUMP FAILED
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
37
NOT REACHED TARGET DRAINAGE AT
T=
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
38
COOLED DRAINAGE FAILED
Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
41
DETERGENT PUMP No.1 FAULT
(flowmeter nr.1 optional)
Make sure that the suction tubes of
can P1 are in a good condition and
not squashed. Repeat the cycle.
Call the Technical Assistance
Service if the fault persists.
42
ACID PUMP No2 FAULT
(flowmeter nr.2 optional)
Make sure that the suction tubes of
can P2 are in a good condition and
not squashed. Repeat the cycle.
Call the Technical Assistance
Service if the fault persists.
43
PUMP 3 INFLOW FAILED
(flowmeter nr.3 optional)
Make sure that the suction tubes of
can P3 are in a good condition and