Config 2 Screen
Figure 7: Configuration 2 page. Note the actual values used may differ from those shown above.
12)
Calibration Gas 1, 2: These boxes are greyed out in this version and the
data is for information purposes only. If you want to change the current
calibration gas then follow the procedure given in section 7.7.
13)
Calibration Constant A & B: These boxes contain the calibration constants
that the Rapidox calculates automatically when you perform a calibration.
You will see that they change each time you perform a new calibration.
14)
O
2
Analogue Outputs (0V/4mA & 10V/20mA): Use these two boxes to set
the High and Low oxygen values for the analogue outputs (6) on the rear
panel. For example if you want 0V/4mA to represent 1% and 10V/20mA to
represent 100% then enter these two values into the appropriate boxes.
The value can be entered either in % or ppm depending on the units
selected. The default values are 100% for the high setting and 0% for the
low setting.
15)
Auxiliary Analogue Outputs (0V/4mA & 10V/20mA): Use these two boxes
to set the High and Low auxiliary values for the analogue outputs (6) on the
rear panel. On the 1100E this is fixed to output pressure from the internal
pressure sensor. For example if you want 0V/4mA to represent -100mbar
and 10V/20mA to represent 100mbar then enter these two values into the
appropriate boxes.