Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
23
Last printed 29/08/2012 16:01:00
21)
Contacts N/O or N/C: This option allows the user to programme how the
relay contacts are set on the rear panel. “Normally Open” means that the
contacts are open circuit when an alarm is not activated and “Normally
Closed” means that the contacts are closed when the alarm is not activated.
22)
Alarm Setpoints: Use this box to programme the set-point value at which
you want the alarm to activate. The units are in %, ppm, or mbar
depending on which sensor you have assigned to the alarm.
Once you have finished editing the parameters, select ‘Write Configuration to
Analyser’ from the file menu or click the ‘Write’ button on any configuration page
and the new data set will be written to the Rapidox. The Rapidox is
reprogrammed using the new configuration. The software then reads the new
settings back from the Rapidox and displays them in the boxes on the left-hand
side confirming that the configuration was successful.
7.6
On-Screen LCD
It is possible to display an LCD emulator on the desktop of your PC. This
reproduces exactly the display on the Rapidox unit and is convenient if you are
some distance away from the machine. To activate the LCD select the ‘Utilities’
menu and select the ‘Show LCD’ option. This will display a small window
showing the LCD that can be positioned independently anywhere on the desktop.
Figure 9: On-screen LCD
The blue LED indicator will flash to show that the display is updating. To close
the LCD either click its window close button, or click the ‘Utilities’ menu on the
main programme window and select the ‘Hide LCD’ option.
7.7
Remote Calibrating
With the on-screen LCD display active it is possible to remotely calibrate the
analyser. Click on the down arrow icon in the bottom right-hand corner of the
LCD display to expand the window: