13
CM-800-001 Rev K
DECEMBER 2019
13
If Dilator 3 was used, it should also be removed once the pedicle
screws are implanted. Adjust the screw head position(s) with
the Screw Head Positioner (CM-131-01) if needed. Orient screw
heads as shown below to ensure easy rod implantation.
ROD INSERTION
DISSECTION
The Ø5.5mm rods are all bullet-nosed to dissect tissue as the
user implants the rod. Prior to inserting the rods, the Dissector
(CM-102-01) can be used to create a pathway through the
tissue below the fascia layer. The Dissector can be inserted
cranially or caudally of the construct but shall follow the
anticipated rod trajectory.