800-572-1028 caltestelectronics.com 2
7. Turn on the high voltage source.
8. You will now be able to analyze the voltage waveform on your
oscilloscope.
9. Turn off the high voltage source.
10.
WARNING
Disconnect the probe from the high voltage source
BEFORE disconnecting the ground clip lead.
3 Frequency Compensation
Proper compensation of the probe is required to assure amplitude
accuracy of the waveform being measured by matching the probe to
the oscilloscope’s input capacitance. Compensation should be adjusted
whenever the probe is connected to, or transferred between, oscilloscopes.
The procedure is the same as for a x10 passive probe.
200 Hz Adjustment
1.
Connect the probe to the oscilloscope.
2.
Apply a 200 Hz square wave to the probe tip and adjust the
oscilloscope controls to display a few cycles of the waveform.
3.
Turn the 200 Hz Trimmer (Figure 2) using the Trimmer Tool,
making a flat topped square wave (Figure 4).
Figure 2: Frequency Compensation Trimmer