12
Ver. 18/12/2020 | Calistair C300
This chapter contains basic safety instructions for handling the decontamination plant
Calistair C300.
It is essential to observe all notes for operation and maintenance of the decontamination plant
Calistair C300 listed in this chapter.
2.1 Personal qualification
The user must have read and understood the operating instructions. The user must correspond
to the target group definition in chapter 1.1.
2.2 Personal protective equipment
Cutting and skin injuries may occur during maintenance measures and repair work.
Therefore wear protective gloves.
2.3 Safety components
Two contact switches between the housing and the housing cover are used for safety shutdown.
Another contact switch under the filter is used to switch off the device.
Specification of the contact switches:
•
Single pole snap switch
•
Normally open contact with quick action contact
•
Plug connection 6.8 x 0.8 mm
•
1E4 T85
•
Switching force 3.5 ± 0.5 N
2. SAFETY
Содержание C300
Страница 2: ......