PGUI / PGUI32 Training Manual

28

1.1

will get an error when you try to execute the transient list. The AC Source does
not automatically switch range to accommodate a higher voltage as this would
cause the output to be dropped temporarily.

8. Starting a transient program with a voltage step is good practice as it will
ensure you start out at this specific voltage value, regardless of what the steady
state value of the source is at the time the step is executed.

9. Move to step No 2  and select ‘V Sweep’ as the transient type. A sweep
allows you to go from the present voltage value to a final voltage value in a
specified period of time.

10. Tab over to the ‘Time  (s)’ field and enter 5.0 for a 5 second sweep. Since
we will sweep from 120 Vrms to 20 Vrms in 5.0 seconds, the Voltage slew rate
will be (120-20)/5 = 20 V/s. There is no need to know this however as the
PGUI will calculate this value for you and program the AC Source transient
system accordingly.

11. Enter 20.0 V for the Voltage value. This will cause the voltage to sweep
down from the 120 V entered in step 1 to 20 volt.

12. Move to the next step (No 3) and select ‘V Srg/Sag’ as the transient type. A
Surge/Sag allows you to go to the specified Voltage value for a period of time
and then return to the previous voltage value.

14. Tab to the ‘Time (s)’ field and enter 20 ms. If our frequency is set to 50 Hz,
this will cause the output to surge to 120 V for one full period of the AC output
and then return back to 20 Vrms.

16. Enter 120.0 for the Voltage value.

17. This is the last event in our sequence so we leave the next step Empty.

18. Your Transient window should look like the window shown here at this
time.

19. If you have unit connected, you can run the transient program by Clicking
on the Start button at the bottom of the Transient Window. If you are in
simulation mode, the program will appear to run the transient and stop after
the 5.04 second duration expires. The Start button is disabled while a transient
program is running.

Next we will perform a similar exercise but this time for the frequency.

Содержание P Series

Страница 1: ...PGUI or PGUI32 TRAINING MANUAL P Series RP Series Graphical User Interface Software 1251RP Programmable AC Source Revision 1 1 Date June 24 1998...

Страница 2: ...PGUI PGUI32 Training Manual 2 1 1...

Страница 3: ...17 3 4 Command Menu 18 3 5 Toolbar 18 3 6 Next 18 4 System Menu 19 4 1 Interface 19 4 2 Configuration Settings 21 4 3 Power On Settings 22 5 Measurements 23 5 1 Enable and Disable Measurements 23 5 2...

Страница 4: ...PGUI PGUI32 Training Manual 4 1 1...

Страница 5: ...a single AC source and will ignore all other types of instruments connected to the same bus It is possible however for other Windows programs to run at the same time as the PGUI and control other inst...

Страница 6: ...Users of Windows NT have to use the 32 bit PGUI32 The 16 bit program will not run under NT PC Based IEEE 488 Interface Controller RP Series AC sources can be controlled remotely using either the IEEE...

Страница 7: ...P Series RP Series 1001P 801RP 1251P 1251RP 2001RP PGUI32 only If you do not have one of these units connected to your PC or have a suitable interface in your PC yet you can still proceed with this t...

Страница 8: ...PGUI PGUI32 Training Manual 8 1 1...

Страница 9: ...ion procedure is very similar except you use the Program Manager s File Run menu to start the SETUP program located on the installation disk 2 1 Windows 95 Installation procedure To install the AC sou...

Страница 10: ...e Programs icon to launch the Windows 95 Add Remove Programs property dialog 4 Click on the Install button located in the top right corner of this dialog box 5 Insert the first installation disk in dr...

Страница 11: ...rectory for the AC source control software which is PGUI on the current drive At this time you can accept the default directory by clicking on the OK button or enter a different directory If the direc...

Страница 12: ...it will be created 11 The setup program will create a program group The default group name used will be PGUI You may also elect to store the AC source control program icon in a different group You ma...

Страница 13: ...disk In Windows 95 the installation program will create a Start menu program group called PGUI and a program shortcut also called PGUI We recommend you keep the distribution disk in a safe place in c...

Страница 14: ...PGUI PGUI32 Training Manual 14 1 1...

Страница 15: ...ry to connect to the AC Source if it was connected at the time the program was closed If not or if the AC source cannot be found at the same IEEE address or serial port the PGUI will launch in simulat...

Страница 16: ...function of the speed at which you drag the slider handle 3 2 By clicking on the arrows located at the left or right end of the control This changes the frequency at 0 1 Hz increments 3 3 By clicking...

Страница 17: ...e can be used to release the unit for front panel access There are situations in which it may be necessary to use the front panel 3 3 Status Bar PGUI status bar Program and interface status PGUI32 sta...

Страница 18: ...th a single mouse click As you get more familiar with operating the PGUI you may rely on the Toolbar more and more as it gets you to the right screen faster Each Toolbar button has a yellow tooltip th...

Страница 19: ...Exercise 1 Pull down the System menu and select the Interface sub menu or click on the IEEE connector symbol in the Toolbar This will bring up the Interface window 2 If you are operating in Simulatio...

Страница 20: ...unicate with the AC source and an error message will appear Note that the P Series does not offer an IEEE 488 interface so you have to use RS232C The RP Series offers both interfaces 6 If you have a u...

Страница 21: ...AC source These settings are determined by the AC Source configuration and cannot be changed by the user Exercise 1 Select the Configuration sub menu from the Source menu 2 Verify that the limits and...

Страница 22: ...front panel setup register 2 as the power on default and click on Close to close the Power on Setting window 3 Next we need to make sure the setup we want the unit to power up in is indeed stored in s...

Страница 23: ...ge measurement accuracy Sudden changes in output voltage will not appear immediately due to this averaging technique The PGUI normally continuously polls the measurement data from the AC Source You ca...

Страница 24: ...xercise we will use training csv 3 Select the Overwrite mode if it is not already selected This will cause a new file to be started each time you start a measurement cycle 4 Select the ASCII string Ti...

Страница 25: ...l or Notepad if Excel is not available to you The following table shows a fraction of the data file 1997 09 17 12 42 23 PGUI Training Manual Date Time F V I 9 17 97 12 42 25 50 271 4 5 9 17 97 12 42 2...

Страница 26: ...PGUI PGUI32 Training Manual 26 1 1...

Страница 27: ...e transients as an exercise Exercise 1 Select the Transient menu which will bring up the Transient window 2 Clear the present contents by clicking on the Clear All key located in the bottom right corn...

Страница 28: ...ent system accordingly 11 Enter 20 0 V for the Voltage value This will cause the voltage to sweep down from the 120 V entered in step 1 to 20 volt 12 Move to the next step No 3 and select V Srg Sag as...

Страница 29: ...nds 3 Set Frequency field to 500 00 Hz This will cause the frequency to sweep up from its present steady state value to 500 Hz 4 Move to the next step and select F Sweep again as the Transient Type Ta...

Страница 30: ...or a 40 msec time Set the Voltage field to 0 0 V Set the Frequency field to 50 00 4 Move to the second row No 2 and select the VF Sweep transient type 5 Set the Time s field to 3 000 for a 3 second sw...

Страница 31: ...e the transient list Time delays can be entered in one of two ways Time in Seconds default mode Time is entered in actual seconds Time in Cycles Time is entered in cycles The length of each cycle is a...

Страница 32: ...ing the last transient program we have created click on the Save button 2 This brings up the File Save dialog box shown here 3 The default directory is the PGUI program directory You can change to any...

Страница 33: ...mpted for a file name If you saved the setup before or you recalled a setup from disk the same setup filename will be used for any subsequent save operations unless you use the Save As sub menu option...

Страница 34: ...settings are different from the one you just saved 2 Pull down the File menu and select the Open sub menu This will show you a list of all setup files in the working directory 3 Select the training s...

Страница 35: ...ed with the program All windows have a Help button which gets you directly to the relevant help topic You can also use the Help menu to see the help table of contents or do a search on a specific topi...

Отзывы: