5
›
english
CaLeXSoFt SoFtWare
SySTEm REquiREmENTS
• CalexSoft software is designed to run under Windows. Suitable versions are Windows
2000, Windows XP and Windows Vista.
• The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
iNSTallaTiON
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
NOTE Do not connect the PyroPen u to the PC before the software is installed
CONNECTiNg ThE PyROPEN u TO a PC
1. Connect the USB Mini B connector on the USB cable provided to the PyroPen U.
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the CalexSoft desktop icon
NOTE Do not disconnect the uSB cable before closing down the CalexSoft
program
ConFiGUration
• Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
• Setting Target Emissivity
To set an Emissivity value for the PyroPen U, use the main program menu to select:
[Setup]
→
[Emissivity]
Emissivity can then be set to a value between 0.3 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
SpeCiFiCation
Temperature Range
-20°C to 500°C
Accuracy
±1% or ±1°C (±2°F) whichever is greater
Repeatability
±0.5% of reading or ±0.5°C (±1°F) whichever is greater
Response Time
500mSec 95% response
Spectral Response
8 to 14µm
Emissivity
adjustable 0.3 to 1.0
Temperature display
0.1°C
Ambient operating range
0 to 50°C (32 to 120°F)
Relative Humidity
10-95% RH non-condensing
Weight
60g
Dimensions
163 x 27 x 16mm
Power
2 x AAA batteries
Laser
Class 2