SERVICE MANUAL
CALIBRATION
3-23
Wait for the phase angle measurement to stabilize on the counter timer. Record the
phase angle as
RDGN4
.
Set the TW frequency to 320 Hz:
SOUR:FREQ 320.0
Wait for the phase angle measurement to stabilize on the counter timer. Record the
phase angle as
RDGN5
.
Set the TW frequency to 60 Hz and output off:
SOUR:FREQ 60.0
OUTP OFF
Calculate the phase angle compensation values:
VALUE1
= -1.0 *
RDGN1
VALUE2
= -1.0 *
RDGN2
VALUE3
= -1.0 *
RDGN3
VALUE4
= -1.0 *
RDGN4
VALUE5
= -1.0 *
RDGN5
Update the phase A offset calibration data via the GPIB:
CAL:UNIQ:OPHA
VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
3.16 PHASE A TO PHASE B CALIBRATION
Note: this calibration step only applies to TW3500 and TW5250 systems.
Connect the counter timer channel 1 to the TW’s A phase through a 10:1 resistor
divider. Connect the counter timer channel 2 to the TW’s B phase through a 10:1
resistor divider. Set the counter time up to measure phase angle.
Set the TW up for 0 degrees phase A to phase B, low range, local sense, 40 Hz, 120V
and close the output by sending the following GPIB commands:
SOUR2:PHAS 0
SOUR:VOLT:RANGE LOW
SOUR:SENSE LOCAL
SOUR:FREQ 40.0
SOUR:VOLT 120.00
OUTPUT ON
Wait for the counter timer to settle and record the phase angle as
RDNG1
.
Содержание TrueWave TW1750
Страница 2: ......
Страница 4: ...ii This page intentionally left blank ...
Страница 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank ...
Страница 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank ...
Страница 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank ...