CALIBRATION
TW SERIES
3-32
Program the TW to 160 Hz using the string below:
SOUR
{1,2,3}
:FREQ 160
Wait until the power analyzer has stabilized and record the current reading as
RDNG3
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS3
.
MEAS
{1,2,3}
:CURR?
Program the TW to 320 Hz using the string below:
SOUR
{1,2,3}
:FREQ 320
Wait until the power analyzer has stabilized and record the current reading as
RDNG4
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS4
.
MEAS
{1,2,3}
:CURR?
Program the TW to 500 Hz using the string below:
SOUR
{1,2,3}
:FREQ 500
Wait until the power analyzer has stabilized and record the current reading as
RDNG5
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS5
.
MEAS
{1,2,3}
:CURR?
Program the TW to 60 Hz and open the output relay using the strings below:
SOUR
{1,2,3}
:FREQ 60
OUTP OFF
Calculate the calibration constants using the following equations:
VALUE1
=
RDGN1
/
MEAS1
VALUE2
=
RDGN2
/
MEAS2
VALUE3
=
RDGN3
/
MEAS3
VALUE4
=
RDGN4
/
MEAS4
VALUE5
=
RDGN5
/
MEAS5
Update the TW’s calibration data:
CAL
{1,2,3}
:MEAS:REM:LOR:CURR:FREQCAL VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
CAL
{1,2,3}
:MEAS:LOC:LOR:CURR:FREQCAL VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
Содержание TrueWave TW1750
Страница 2: ......
Страница 4: ...ii This page intentionally left blank ...
Страница 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank ...
Страница 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank ...
Страница 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank ...