MN220-1 B 48
Portable Oxygen Concentrator
1.
Turn unit off and disconnect the power cord.
2.
Remove the back housing, and locate the product filter/volume tank. See figure below.
3.
Disconnect tubing from the inlet of the product filter/volume tank.
4.
Unscrew the product filter/volume tank from the product manifold
5.
Place the o-ring into the o-ring groove located on top of the volume tank.
6.
Fasten the new product filter/volume tank to the product manifold.
7.
Connect the tubing and secure with a tie-wrap.
Figure 9: Product Filter
If the compressor inlet felt filter (shown above) is contaminated, a
replacement felt filter can be ordered by contacting CAIRE and
referencing Part Number
FI143-2
INFECTION CONTROL
With growing concern about possible cross infection from home oxygen equipment from one
home care patient to another, a clarification on this topic is necessary. The organisms of most
concern are M. Tuberculosis, HIV, and Viral Hepatitis. These are potentially pathogenic.
Product
Filter
Compressor
Felt Filter