B-701964-00 Rev. C
Service and Repair
5-15
Companion 1000/T Liquid Oxygen System Technical Manual
5.10
FEMALE FILL CONNECTOR
The female fill connector assembly is a special valved coupling that enables liquid
oxygen to transfer from a Companion Stationary unit to the Companion 1000/T
Portable. An internal cartridge assembly with a spring-loaded Kel-F poppet creates a
pressure-tight seal when the female fill connector is not coupled to a Stationary unit.
During a fill, a spring-energized Teflon lip seal prevents leakage between the female fill
connector and the Stationary unit male fill connector.
5.10.1
Removal
Refer to Figures 6-1 or 6-2 and 6-8 to identify parts with bold number references.
1. Remove side cover "B" (Section 5.2.1, steps 1 and 2).
2. Use a
9
/
16
-in. open-end wrench to disconnect the tube nut
7
at the female adaptor
4
(Figure 5-19).
Figure 5-19
3. Use a
7
/
8
-in. open-end wrench to loosen the jam nut
5
(earlier models may have a
5
/
8
-in. jam nut) that holds the female adapter to the mounting bracket. Pull the fill
connector assembly
1
down slightly to clear the fill tube
8
and then slide it out of
the mounting bracket (Figure 5-20).
Figure 5-20
Remember to always use a backup wrench when loosening or tightening
connections.
Содержание COMPANION 1000
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 10: ...Companion 1000 T Liquid Oxygen System Technical Manual vi B 701964 00 Rev C This page intentionally left blank...
Страница 73: ...2EV 6 1 Section 6 PARTS LISTS AND ILLUSTRATIONS...
Страница 85: ......
Страница 86: ...20 2 Inc All rights reserved B 701964 00 Rev 6 RG MVTSVX RHYWXVMEP V 7XI EPP VSYRH USA Toll Free 1 800 Rx ONLY...