Performance Verification
B-701964-00 Rev. C
Perform the following Liquid Leak Detector Test (Section 3.2.1) to determine if there
are any substantial leaks. After performing this procedure, if you feel there is an
unacceptable amount of leakage, perform the Pressure Hold Test (Section 3.2.2) to
determine if the total leak rate is acceptable.
3.2.1
Liquid Leak Detector Test
PROCEDURE
1. Set the flow control valve to 0 L/min (off).
2.
Warm, Empty Unit:
Pressurize the Companion 1000/T Portable to approximately
22 psig (152 kPa) with gaseous oxygen using the portable pressurizing setup and
gaseous oxygen supply (Figure 3-1). Close the vent valve after pressurizing the unit
and remove the pressurizing setup from the vent tube.
Figure 3-1: Pressurizing the Companion 1000/T through the Vent Circuit
Connect the test pressure gauge to the oxygen outlet barbed fitting and set the flow
control valve to the highest setting. Place the Companion 1000/T in the portable test
fixture (Figure 3-2).
You may perform the Liquid Leak Detector Test on units that contain liquid
and/or gaseous oxygen.
Test Pressure
Gauge
P/N B-775270-00
Tubing
P/N B-778214-00
Vent Tube
Vent Valve Open
Clamp
Tee
P/N B-778211-00
Adjustable 0-50 psig (0-345 kPa)
Gaseous Oxygen Source
Portable
Test Fixture
P/N B-778202-00
Содержание COMPANION 1000
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 10: ...Companion 1000 T Liquid Oxygen System Technical Manual vi B 701964 00 Rev C This page intentionally left blank...
Страница 73: ...2EV 6 1 Section 6 PARTS LISTS AND ILLUSTRATIONS...
Страница 85: ......
Страница 86: ...20 2 Inc All rights reserved B 701964 00 Rev 6 RG MVTSVX RHYWXVMEP V 7XI EPP VSYRH USA Toll Free 1 800 Rx ONLY...