![CAENels Flucs CT -100 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/caenels/flucs-ct-100/flucs-ct-100_user-manual_3283200023.webp)
CT-100/CT-150
Series User’s Manual
23
Secondary Coil Resistance -
R
SEC
43
Ω
34
Ω
Accuracy (typ.)
*
< 30 ppm / FS
< 0.25% / FS
(“V”
- version)
Connections
D-sub 9 connector ("C"-model)
7-pin strip TH type ("P"-model)
Operating Temperature
Range
0°C
–
50°C
Mechanical (Outer)
Dimensions
45 x 57 x 75 mm
Primary Conductor
Hole Diameter - Ø
16 mm
Maximum Weight
250 g
* These specifications are guaranteed only when the CT sensor is used with the dedicated CAEN ELS PS1215 power supply
Table 4:
Technical Specification
4.1
Equivalent Input Noise
The typical equivalent input noise of the CT-100/CT-150 transducers is
hereafter presented as a function of the measuring bandwidth and of the output version.
The connector topology does not affect the noise performances of the devices and the
characterization is performed on the “C” connector versions. The table and its graph
are valid for both current and voltage output options.
CT-100/CT-150 Typical Equivalent Input Noise
Bandwidth
CT-100
CT-150
200 Hz
0.5
1.6
1 kHz
2.5
3.7
10 kHz
3.5
6.3
50 kHz
4.5
6.8
Table 5:
Equivalent Input Noise (typical values)
Содержание Flucs CT -100
Страница 28: ...DC User s Manual...