![Caen V1729 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/caen/v1729/v1729_technical-information-manual_3283096011.webp)
PRELIMINARY
Document type:
Title:
Revision date:
Revision:
User's Manual (MUT)
Mod. V1729 4 Channel 12 Bit Sampling ADC
22/06/2005
3
NPO:
Filename:
Number of pages:
Page:
00109/04:V1729.MUTx/03 V1729_REV3.DOC
38
11
The user if necessary sends a software trigger (in the case of an automatic trigger)
having taken care to wait at least the time necessary for the PRETRIG.
- The program then waits for a request emitted by the V1729 board when the data are
ready to be read. For this, there exist two possibilities illustrated by Fig 2.5 :
Waiting and handling of an interruption (SRQ in GPIB, IRQ3 in VME).
Regular scanning of the LSB of the interruption register (INTERRUPT h80).
In both cases, the user should acknowledge the request by writing a 0 in the
INTERRUPT
register, but the latter is anyhow reset by the START_ACQ command. The
first solution has the big advantage of being less noisy for the front-end of the board.
- The user can then partially or wholly read the data stored in the board. This includes :
the values of the vernier and of the samples (see 4.7 )
the value of TRIG_REC which permits determination of the position of the trigger in the
acquisition window (see 2.5.2)
- The user must correct the data before using it (see 2.5):
By subtracting the pedestals (see 2.1.1).
By reordering the data (see 2.5.2).
These operations can be executed on-line or off-line according to the context.
2.5 Correction of data
The raw data extracted from the board must be treated before really being usable.
This treatment consists of two operations :
Correction of the pedestals.
Reordering and time alignment of the data.