Document type:
Title:
Revision date:
Revision:
User's Manual (MUT)
Mod. N957 8k Multi-Channel Analyzer
28/05/2012
6
NPO:
Filename:
Number of pages:
Page:
00105/04:N957x.MUTx/06 N957_REV6
35
34
Fig. 5.4: Histogram plot of
60
Co source in progress
Type ‘r’ key to reset the histogram and the ‘p’ key to pause/resume the
acquisition.
5. Type ‘s’ key to stop the acquisition session:
Fig. 5.5: Acquisition report
The acquisition report is displayed, with the number of the input signal samples
(ADC Conversions Number) and the contents of the two board’s 32-bit timers
(Timer and Live Time).
6. Type any key to quit the program.
Raw data (i.e. the samples of the digitized input signal) are saved into the
output.log
file,
according to the configuration file. The histogram data are saved into the
hist.dat
file.