Interlock signal 

With reference to diagram of  the Interlock input shown above, the diode is part of opto-coupler stage. 
Interlock means that channels are hardware disabled. The interlock operation is as follows: 

CONFIGURATION  

INTERLOCK STATUS 

leave contact open 

INTERLOCK 

voltage level (0÷1V, ~5mA current) between pin 2 and pin 3 

INTERLOCK 

short circuit pin 1 with pin 2, and pin 3  with pin 4 

ENABLED 

voltage level (4÷6V, ~5mA current) between pin 2 and pin 3 

ENABLED 

The front panel Interlock LED is ON when the INTERLOCK is active;  as INTERLOCK is active, channels are 
turned off at the fastest available rate, regardless the RAMP DOWN setting. 

DC Input  

 

Fig. 9: 12V DC Input 

NAME: 

TYPE: 

SIGNAL: 

FUNCTION: 

12V In 

RAPC722X SWITCHCRAFT PCB 2.1mm DC Power Jack  

Input 

+12V DC Input 

ON/OFF switch 

Molveno A11331122000 A1 switch 

 

 

power supply OFF.

 

I  

→ 

power supply ON 

IMON Range 

The channel current can be monitored with an increased resolution (10x) in the following ranges 

(“low” 

range): 

DT5519E 

 300 µA 

DT5521E 

 30 µA 

DT5521HE 

 2 µA  

DT5533E 

 300 µA 

DT5534E 

 100 µA 

By selecting IMON Range = LOW the output current is monitored with 

DT5519E 

5 nA resolution (instead of 50 nA), in the 0 

 300 µA range 

DT5521E 

500 pA resolution (instead of 5 nA), in the 0 

 30 µA range 

DT5521HE 

100 pA resolution (instead of 1 nA), in the 0 

 2 µA range 

DT5533E 

5 nA resolution (instead of 50 nA), in the 0 - 300 µA range 

DT5534E 

2 nA resolution (instead of 20 nA), in the 0 - 100 µA range 

It is important to notice that, if IMON Range = LOW (see p.15) is selected, and the channel draws a current 
larger than  

300 µA 

DT5519E 

then Overcurrent is signalled.

 

30 µA 

DT5521E 

2 µA  

DT5521HE 

300 µA 

DT5533E 

100 µA 

DT5534E 

Содержание DT5519EM

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Es DT55xxE Desktop HV Power Supply Rev 6 11 April 2017...

Страница 2: ...roduct is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsi...

Страница 3: ...r 11 Power Supply Modules LabVIEW Instrument Driver 11 4 VT emulator control 12 Module access 12 Ethernet configuration 12 Firmware upgrade 13 Format EEPROM 14 Channels settings 14 Communication Proto...

Страница 4: ...EP 6 kV 1 mA 4 W Positive HV outputs are delivered through SHV connectors Mixed versions have two positive and two negative channels The HV output RAMP UP and RAMP DOWN rates may be selected independe...

Страница 5: ...cable 10BASE T Ethernet cable Power requirements The module is powered by the external AC DC stabilized power supply included in the delivered kit Switchbox Model FRA045 S12 4 see specifications in t...

Страница 6: ...3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated pairs Current rating 10 A WARNING These connectors produce extremely hazardous high voltages at a potential...

Страница 7: ...trol panel NAME TYPE SIGNAL FUNCTION ALARM RED LED AMP 280371 2 Out Alarm status signalled active LOW INTERLOCK RED LED AMP 280371 2 In Interlock signal channels hardware disabled Alarm signal As an A...

Страница 8: ...FUNCTION 12V In RAPC722X SWITCHCRAFT PCB 2 1mm DC Power Jack Input 12V DC Input ON OFF switch Molveno A11331122000 A1 switch O power supply OFF I power supply ON IMON Range The channel current can be...

Страница 9: ...nA 2 of read 5nA 2 of read 500pA 2 of read 100nA 2 of read 50nA ISET vs IMON High range 2 of read 1 A 2 of read 50nA 2 of read 5nA 2 of read 1 A 2 of read 500nA Low range 2 of read 100nA 2 of read 5nA...

Страница 10: ...arts DT55xxE desktop power supply unit External AC DC power supply with separate power cord USB cable 10BASE T Ethernet cable Moreover an external PC is required supported OS Windows 7 or later Linux...

Страница 11: ...r level application such as GECO2020 and the lower layer software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Pow...

Страница 12: ...k and set the virtual communication port associated to the module Select USB connection and the used port number set port as follows baud rate 9600 Data bits 8 Parity none stop bit 1 Flow control none...

Страница 13: ...in the EEPROM and go back to Main Menu the DHCP server will automatically assign a new IP to the module at next Power On Firmware upgrade To upgrade the firmware download from www caen it unit page th...

Страница 14: ...ded firmware Format EEPROM By typing F on Terminal Board Menu Fig 10 it is possible to access the format EEPROM menu Fig 12 Format EEPROM Menu Channels settings By typing D on Terminal Board Menu Fig...

Страница 15: ...0x1000 At least one ch in Interlock CMD MON PAR MACADDR Board MAC Address CMD MON PAR IPADDR Board IP Address CMD MON PAR SUBMASK Subnet mask CMD MON PAR GATEWAY Gateway CMD MON PAR DHCPEN DHCP Enable...

Страница 16: ...hannel in Temperature warning 80 C IS_OVT 0x400 Channel in OVT 125 C IS_KILL 0x800 Channel Killed IS_INTLCK 0x1000 Channel in Interlock CMD MON CH XX PAR RUP RAMP UP value CMD MON CH XX PAR RUPMIN RAM...

Страница 17: ...nge of customers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Ve...

Отзывы: