background image

CAEN

 

 

Electronic Instrumentation

 

 

DT55xxE Desktop HV Power Supply 

Technical specifications table 

Table 1: DT55xxE Series Channel technical specifications 

Model 

DT5519E 

DT5521E 

DT5521EH 

DT5533E 

DT5534E 

Packaging 

Desktop module housed in a 154x50x164 mm

alloy box; weight:~800g 

Output channels 

Positive, Negative and Mixed (2+2) Polarity available 

Output ranges 

500 V/3mA 

6kV/300 µA 

6kV/20 µA 

4kV/3mA 

6kV /1mA 

Max. Ch. Output Power 

1.5 W 

1.8 W 

180 mW 

4W 

Vset Resolution 

10 mV 

100 mV 

Vmon Resolution 

1 mV 

50 mV 

10 mV 

50 mV 

ISET Resolution 

50nA 

5 nA 

500 pA 

50 nA 

20 nA 

IMON Resolution 

High range 

10nA 

1 nA 

100 pA 

10 nA 

5 nA 

Low range

1

 

1nA 

100 pA 

10 pA 

1 nA 

500 pA 

VMAX hardware 

0 ÷ 510 V 

0 ÷ 6100 V 

0 ÷ 4100 V 

0 ÷ 6100 V 

VMAX hardware resolution 

1 V 

VMAX hardware accuracy 

 

± 2% of FSR 

IMAX 

3100 µA 

310 µA 

21 µA 

3100 µA 

1050 µA 

Alarm output 

Open collector, 100 mA maximum sink current 

Interlock input 

LOW: <1V; current~5mA; HIGH: 4÷6 V 

Ramp Up/Down 

1÷100 V/s, 1 V/s step 

1÷500 Volt/s, 1 Volt/s step 

A

ccur

acy

2

 

Vmon vs. Vout 

±0.05% of read  ±0.5V 

±0.05% of read  ±1V  ±0.05% of read  ±1V 

±0.05% of read  ±1V 

±0.05% of read  ±1V 

Vset vs. Vout 

±0.05% of read  ±0.5V 

±0.05% of read  ±1V  ±0.05% of read  ±1V 

±0.05% of read  ±1V 

±0.05% of read  ±1V 

IMON  vs. 
Iout 

High range 

±2% of read ±1µA 

±2% of read  ±50nA 

±2% of read  ±5nA 

±2% of read ±1µA 

±2% of read ±500nA 

Low range 

±2% of read ±100nA 

±2% of read  ±5nA 

±2% of read  ±500pA 

±2% of read ±100nA 

±2% of read  ±50nA 

ISET  vs. 
IMON 

High range 

±2% of read ±1µA 

±2% of read  ±50nA 

±2% of read  ±5nA 

±2% of read ±1µA 

±2% of read ±500nA 

Low range 

±2% of read ±100nA 

±2% of read  ±5nA 

±2% of read  ±500pA 

±2% of read ±100nA 

±2% of read  ±50nA 

Voltage Ripple

3

 

 

typical 

<3 mVpp 

<12 mVpp 

<15 mVpp 

max 

<5 mVpp  

<20 mVpp  

<25 mVpp  

Ventilation  Fans 

Two Sunon MC35101V1-000U-A99 

Humidity range 

0 ÷ 80% 

Operating temperature 

0 ÷ 45°C 

Storage temperature 

-10 ÷ 70°C 

Vout / Temperature  coefficient 

max. 50ppm / °C 

IMON / Temperature  coefficient 

max 100ppm/C° 

Long term stability Vout vs. Vset 

± 0.02% (after one week  @ constant temperature) 

External power supply

 

Switchbox FRA045-S12-4 (12 VDC, 3.75 A, 45 W); Universal Input C14 receptacle;  
INPUT Voltage range 100-240VAC 1.2A 50-60Hz. Inrush current 40A at 115VAC / 80A at 230VAC max. Dielectric 
withstand Input/output 3,000VDC.  
OUTPUT Output v12V. Ripple and noise 2% p-p max. Load regulation ±5% max. No load stand by power <0.5W 
@ 230VAC. Efficiency >=85% for CEC requirement. Hold up time 10ms at nominal line. Protections OCP, OVP, over 
power & short circuit.  
GENERAL Std output connector Dc barrel jack. Std output cable/length UL1185, #18AWG / 5 ft.  
ENVIRONMENTAL Operating temperature 0°C to +40°C. Storage temperature 

20°C to +85°C.  

STANDARDS Safety standards IEC/UL/EN60950-1, CE, CB. EMC EN55022 (CISPR 22) class B, FCC class B. 

 

                                                                        
 
 

1

 Available as optional customization 

2

 Accuracy values are measured from 10% to 90% of Full Scale Range 

3

 Measured with: 1m cable length; 2nF capacitance, 100MHz band width 

Содержание DT5519EM

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Es DT55xxE Desktop HV Power Supply Rev 6 11 April 2017...

Страница 2: ...roduct is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsi...

Страница 3: ...r 11 Power Supply Modules LabVIEW Instrument Driver 11 4 VT emulator control 12 Module access 12 Ethernet configuration 12 Firmware upgrade 13 Format EEPROM 14 Channels settings 14 Communication Proto...

Страница 4: ...EP 6 kV 1 mA 4 W Positive HV outputs are delivered through SHV connectors Mixed versions have two positive and two negative channels The HV output RAMP UP and RAMP DOWN rates may be selected independe...

Страница 5: ...cable 10BASE T Ethernet cable Power requirements The module is powered by the external AC DC stabilized power supply included in the delivered kit Switchbox Model FRA045 S12 4 see specifications in t...

Страница 6: ...3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated pairs Current rating 10 A WARNING These connectors produce extremely hazardous high voltages at a potential...

Страница 7: ...trol panel NAME TYPE SIGNAL FUNCTION ALARM RED LED AMP 280371 2 Out Alarm status signalled active LOW INTERLOCK RED LED AMP 280371 2 In Interlock signal channels hardware disabled Alarm signal As an A...

Страница 8: ...FUNCTION 12V In RAPC722X SWITCHCRAFT PCB 2 1mm DC Power Jack Input 12V DC Input ON OFF switch Molveno A11331122000 A1 switch O power supply OFF I power supply ON IMON Range The channel current can be...

Страница 9: ...nA 2 of read 5nA 2 of read 500pA 2 of read 100nA 2 of read 50nA ISET vs IMON High range 2 of read 1 A 2 of read 50nA 2 of read 5nA 2 of read 1 A 2 of read 500nA Low range 2 of read 100nA 2 of read 5nA...

Страница 10: ...arts DT55xxE desktop power supply unit External AC DC power supply with separate power cord USB cable 10BASE T Ethernet cable Moreover an external PC is required supported OS Windows 7 or later Linux...

Страница 11: ...r level application such as GECO2020 and the lower layer software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Pow...

Страница 12: ...k and set the virtual communication port associated to the module Select USB connection and the used port number set port as follows baud rate 9600 Data bits 8 Parity none stop bit 1 Flow control none...

Страница 13: ...in the EEPROM and go back to Main Menu the DHCP server will automatically assign a new IP to the module at next Power On Firmware upgrade To upgrade the firmware download from www caen it unit page th...

Страница 14: ...ded firmware Format EEPROM By typing F on Terminal Board Menu Fig 10 it is possible to access the format EEPROM menu Fig 12 Format EEPROM Menu Channels settings By typing D on Terminal Board Menu Fig...

Страница 15: ...0x1000 At least one ch in Interlock CMD MON PAR MACADDR Board MAC Address CMD MON PAR IPADDR Board IP Address CMD MON PAR SUBMASK Subnet mask CMD MON PAR GATEWAY Gateway CMD MON PAR DHCPEN DHCP Enable...

Страница 16: ...hannel in Temperature warning 80 C IS_OVT 0x400 Channel in OVT 125 C IS_KILL 0x800 Channel Killed IS_INTLCK 0x1000 Channel in Interlock CMD MON CH XX PAR RUP RAMP UP value CMD MON CH XX PAR RUPMIN RAM...

Страница 17: ...nge of customers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Ve...

Отзывы: