EN 12830 Compliance Matrix
§
Requirement
Compliance
Note
4.1
The means used by the themometer must be independent of
any temperature measurement which is used to control the
refrigeration system
Yes
Independent stand alone self
powered loggers
4.2
The measurement range of the sensors must be <-25 °C >15 °C
with minimum span of 50°C
Yes
Temperature measurement
operating range is -30°C to +70°C
4.3
The time and date of the beginning shall be readable and any
settings which configure the recording shall be protected against
un-authorised or accidental modification
Yes
All data inside memory can be
password protected according to
EPC Class1 Gen2 standard
4.4.1
At least the time and temperature shall be recorded. The place
of measurement has to be indicated
Yes
Time and temperature are
recorded.
The user memory of the tag (54
byte) can be used to store the
place of measurement
4.4.2
It shall be possible to identify and consult the charts and the
recorded data. Records shall be available for at least one year
Yes
All records are identified and
stored on the users system
indefinitely
4.5
Devices with an autonomous power supply shall be marked as
such with an indication of the temperature range. An indication
of battery life is desirable
Yes
The recommended operating
range is reported on the user
manual. The battery charge status
can be detected through RFID
interface
4.6
The degree of protection shall be at least IP20, IP55 or IP65 for
different environments
Yes
Degree of protection is IP 67
5
4.8.5
Electrical power disturbances and susceptible to radiated
electromagnetic field
Yes
See test report
6
4.9.2.1
Under rated operating conditions the recorder shall have a
maximum error of ±1°C and a resolution of better than ±0.5°C
Yes
Resolution is 0.1°C.
Temperature accuracy is +/-0.1°C
(typical), +/-0.75°C (worst case)
7
4.9.2.2
The maximum recording interval for transport shall be 5m for
recording durations shorter than 24 Hrs and 15m for 1-7 days
and 60m for durations longer than 7 days. For storage, the
recording interval shall be 60m
Yes
The recording interval is fully
programmable from 8s to 18hours
4.9.2.3
The manufacturer shall specify the recording capacity as either
duration at a given interval or as a number of readings
Yes
The recording capacity is 8k
samples (temperature only) or 2k
samples (temperature and
timestamp)
4.9.2.4
The maximum relative timing error shall be better that 0.2% for
up to 31 days and 0.1% for durations > 31 days
Yes
The maximum timing error is
0.01%
3
4.9.2.5
Response Time shall have a maximum response time of 60 min.
(90%)
Yes
The response time is 20min.
3
4.9.3.1
Tab
3D
The standard operating range is -30 to +70 °C operationally and -
40 to +85 °C for storage
Yes
See test report
3
4.9.3.2
The equipment shall withstand vibration from 5Hz to 8.6 Hz
10mm amplitude and 8.6Hz to 150 Hz acceleration of 3g
Yes
See test report
8
4.9.3.3
The equipment shall withstand shocks with the following
characteristics: acceleration 10g, duration 10ms, number of
shocks 1
Yes
See test report
4
4.10
The data shall be protected against alteration
Yes
All data inside memory can be
password protected according to
EPC Class1 Gen2 standard
5
TesLab s.r.l. - Environmental Test Report – A927Z UHF Semi-Passive Logger Tag - Ref.TesLab 09B204A
6
GSD s.r.l. – CE Test Report - A927Z UHF Semi-Passive Logger Tag - Ref. 29346A
7
CAEN RFID s.r.l. internal test report: A927Z EN 12830 Test Report – 01/03/2010
8
GSD s.r.l. – Vibration Test Report – A927Z UHF Semi-Passive Logger Tag - Ref. 29340