A927ZET Technical Information Manual
27
Precision/Data Type
Precision represents the sensor accuracy expressed in tenth of the sensor base unit:
Temperature: 0.1 °C
Battery: 0.1 V
Temperature Precision/Data Type (int) [15:00]
User
Bank:11b
Word addr: Ch
R only
Temperature Precision/Data Type (ext) [15:00]
User
Bank:11b
Word addr: 101Ch
R only
Battery Precision/Data Type [15:00]
User
Bank:11b
Word addr: 202Ch
R only
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Data Type [7:0]
Precision [7:0]
Temperature
Battery
Precision [7:0]
1
1
Data Type [7:0]
RFU
RFU
Offset
To obtain the effective value of the physical parameter, the Offset Value has to be added to the Sensor data value
(stored in the sensor data logging area).
Temperature Offset (int) [15:00]
User
Bank:11b
Word addr: Dh
R only
Temperature Offset (ext) [15:00]
User
Bank:11b
Word addr: 101Dh
R only
Battery Offset [15:00]
User
Bank:11b
Word addr: 202Dh
R only
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Offset [15:00]
Temperature
Battery
Offset [15:00]
0
0
Low Limit
Low Limit is the minimum measurable value of applied parameter. It is expressed in twos complement format.
Temperature Low Limit (int) [15:00]
User
Bank:11b
Word addr: Eh
R only
Temperature Low Limit (ext) [15:00]
User
Bank:11b
Word addr: 101Eh
R only
Battery Low Limit [15:00]
User
Bank:11b
Word addr: 202Eh
R only
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Low Limit [15:00]
Temperature
Battery
Low Limit [15:00]
FFECh (-20°C)
2 (2V)