27
EN
NHM
≤
250
μ
S/cm
TH < 8 °dH, Cl-
≤
25 ppm (5 drops of
reagent*)
No water treatment
TH
≥
8 °dH, Cl-
≤
25 ppm (5 drops of
reagent*)
Cadco
.Finest
Cl- > 25 ppm (5 drops of reagent*)
Reverse Osmosi System
NHM = 250 - 350
μ
S/cm
Cl-
≤
20 ppm (4 drops of reagent*)
Cadco
.Finest
Cl- > 20 ppm (4 drops of reagent*)
Reverse Osmosi System
NHM = 350 – 450
μ
S/cm
Cl-
≤
15 ppm (3 drops of reagent*)
Cadco
.Finest
Cl- > 15 ppm (3 drops of reagent*)
Reverse Osmosi System
NHM = 450 – 500
μ
S/cm
Cl-
≤
10 ppm (2 drops of reagent*)
Cadco
.Finest
Cl- > 10 ppm (2 drops of reagent*)
Reverse Osmosi System
NHM
≥
500
μ
S/cm
Reverse Osmosi System
* use the reagent in the
Cadco
KSTR1081A kit
** hardness
measured
using
the
Cadco
KSTR1082A kit
WATER SPECIFICATIONS FOR THE STEAM CIRCUIT (STEAM INSIDE THE
OVEN CAVITY)
Free chlorine
≤
0.1 ppm, Chloramines
≤
0.1 ppm (mg/L)
TDS
≤
125 ppm, Silica
≤
12 ppm, pH 7 to 8.5, Total hardness (TH)
≤
8 °d
Chlorides
≤
25 ppm, Alkalinity
≤
150 ppm as CaCO
3
The maximum chloride content, Cl-, is not a fixed value; it depends on the
calculated Non Hardness Measurement (NHM)
NHM = EC - (TH** x 30)
EC = electrical conductivity [
μ
S/cm]
TH = total hardness** [° dH - German degrees]
Содержание BAKERLUX SHOP.Pro XAKHT-HCFS
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 15: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 21: ......
Страница 22: ......
Страница 29: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 37: ......
Страница 38: ......
Страница 40: ......