13
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoid the liquid penetration inside the
device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the device to CA-MI
technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information that can
help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI
TECHNICAL SERVICE.
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE
TAMPERED.
INSTRUCTIONS
•
Place the unit on a flat, horizontal surface.
•
Connect the short silicon tube
3
, with antibacterial filter
1
, to the suction connector
4
. The other tube, with one end connected to the antibacterial
filter with the other end to jar’s lid connector
2
where has been fixed the red float. When the 90% of the volume of the jar is reached there is the
activation of the security float (the float close the aspiration connector on the jar) to the avoid liquid penetration inside the device.
•
Connect the long silicon tube
5
to the other jar's lid connector
6
•
Connect the other end of the long silicon tube
5
to the probe plastic connector
8
then connect the suction probe to it.
•
Connect the power cord to the device then connect the plug to the electrical mains supply.
•
Push switch
7
on position
I
to start suction
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the functionally vacuum)
then rescrew the lid on the jar correctly.
•
Once finished push switch on
0
position and unplug.
•
To extract the accessories and start with cleaning.
WARNING:
Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the “VACUUM”.
A wrong connection causes immediate destruction in case of contact with sucked liquids.
Battery functions
The internal battery provides 45 to 50 minutes of continuous use.
After use, the unit must be recharged for 120 to 150 minutes. A standard cycle would be to use the equipment for 45 minutes and then recharge it for 120
minutes. A complete recharge takes 8 hours.
To recharge, connect the power cable to a 230V outlet and ensure the power switch on the unit is turned on.
Battery function is indicated by the LED lights at the top right of the unit.
The green LED is on if either the battery is fully charged or the mains power has been connected. The yellow LED indicates the unit is charging.
The red LED indicates the battery is fully discharged.
NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT SURVEILLANCE
ALWAYS PLACE THE DEVICE IN POSITIONS FOR EASY DISCONNECTION