Guidance and manufacturer’s declaration – Electromagnetic Emissions 

The surgical aspirator 

NEW ASKIR 230/12V BR 

is intended for use in the electromagnetic environment specified below. The customers or the user 

of the surgical aspirator 

NEW ASKIR 230/12V BR 

should assure that it’s used in such an environment. 

Emissions Test 

Compliance 

Electromagnetic environment - guidance 

Irradiated / Conducted emissions CISPR11 

Group 1 

The surgical aspirator NEW ASKIR 230/12V BR only used RF energy 
only for its internal functioning. Therefore its RF emissions are very low 
and are not cause interference in proximity of any Electronic appliances. 

Irradiated / Conducted emissions CISPR11 

Class [B] 

The surgical aspirator NEW ASKIR 230/12V BR can be used in all 
environments, including domestic and those connected directly to the 
public mains distribution that supplies power to environments used for 
domestic scopes or environments feeds to you from batteries. 

Harmonic emissions IEC/EN 61000-3-2 

Class [A] 

Voltage fluctuations / flicker emissions  
IEC/EN 61000-3-3 

Complies 

 
 

Guidance and manufacturer’s declaration – Electromagnetic Emissions 

The surgical aspirator 

NEW ASKIR 230/12V BR 

is intended for use in the electromagnetic environment specified below.  

The customers or the user of the surgical aspirator  

NEW ASKIR 230/12V BR

 should assure that it’s used in such an environment. 

Immunity Test 

Compliance 

Electromagnetic environments - guidance 

Electrostatic discharge (ESD) 
IEC/EN 61000-4-2 

± 6kV on contact 

± 8kV in air 

Floors should be wood, conceret or ceramic tile. If floors are covered with 
synthetic material, the relative humidity should be at least 30%. 

Electrical fast transient / burst 
IEC/EN 61000-4-4 

± 2kV power supply 

Mains power quality should be that of a typical commercial environment 
or hospital or environments feeds to you from batteries. 

Surge  
IEC/EN 61000-4-5 

± 1kV differential mode 

Mains power quality should be that of a typical commercial environment 
or hospital or environments feeds to you from batteries. 

Loss of voltage, brief voltage 
interruptions and variations 
IEC/EN 61000-4-11 

5%U

T

 for 0.5 cycle 

40%U

T

 for 05 cycle 

70%U

T

 for 25 cycle 

<5%U

T

 for 5 sec 

Mains power quality should be that of a typical commercial environment 
or hospital If the user of the surgical aspirator NEW ASKIR 230/12V BR 
request that the appliance operates continuosly, the use of a continuity 
unit is recommended. 

Magnetic field 
IEC/EN 61000-4-8 

3A/m 

The power frequency magnetic field should be measured in the intended 
installation location to assure that it’s sufficiently low. 

Conducted Immunity 
IEC/EN 61000-4-6 

3Vrms 150kHz to 80MHz  

(for appliances that aren’t 

life - supporting) 

Irradiated Conducted 
IEC/EN 61000-4-3 

3V/m 80MHz to 2.5 GHz 

(for appliances that aren’t 

life - equipment) 

Note U

T

 is the value of the power supply voltage 

 

 

CLEANING OF ACCESSORIES 

 
To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions. 
Washing and / or cleaning the autoclavable jar as to be carried out as follows: 

 

Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances; 

 

Disconnect the jar from the device 

 

Disconnect all tubes from the jar and the protection filter 

 

Empty and dispose of the content and of the suction catheter according to the laws in force in your country; 

 

Separate all parts of the cover (overflow valve, o-ring); 

After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.                                                            
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to remove incrustations. 
Rinse in running warm water and dry all parts with a soft cloth (non-abrasive). 
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure) 
making sure that the jar is positioned upsidedown. 
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1).                             
Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended. 
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged. 
Assemble the jar as follows: 

 

Place the overflow valve into its seat in the cover (under VACUUM connector) 

 

Insert floating valve keeping the o-ring towards the opening of the cage 

 

Place the o-ring into its seat around the cover 

 

After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid exit 

 

The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C. The conical connector can be sterilized on autoclave using a 
sterilization cycle at 121°C.  The device is ready for a new employment now.  
 

 

DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER

 

 

Содержание NEW ASKIR 230/12V BR

Страница 1: ...12V BR NEW ASKIR 230 12V BR ASP RAT R NEW ASKIR 230 12V BR MANUALE D USO INSTRUCTION MANUAL MODE D EMPLOI HANDBUCH MANUAL DE ISTRUCCIONES KULLANIM KILAVUZU CA MI Srl Via Ugo La Malfa nr 31 43010 Pilas...

Страница 2: ...2...

Страница 3: ...iane e stabili Posizionare l apparecchio in modo da evitare di occluderne le prese d aria poste sul retro Non utilizzare l apparecchio in ambienti in cui presenti miscele anestetiche infiammabili con...

Страница 4: ...ASSORBITA 50VA FUSIBILE F 1 x 630mA 250V ASPIRAZIONE MASSIMA senza connessione vasi 75kPa 0 75 Bar ASPIRAZIONE MINIMA senza connessione vasi Minore di 40kPa 0 40 bar FLUSSO MASSIMO D ASPIRAZIONE senza...

Страница 5: ...te commerciale o ospedale o ambienti alimentati da batterie Buchi di tensione brevi interruzioni e variazioni di tensione IEC EN 61000 4 11 5 UT for 0 5 cycle 40 UT for 05 cycle 70 UT for 25 cycle 5 U...

Страница 6: ...e poter effettuare un ciclo di sterilizzazione ad una temperatura di 120 C Il raccordo conico che viene fornito insieme ai tubi di aspirazione pu essere sterilizzato ad una temperatura di 121 C NON LA...

Страница 7: ...are a fondo il coperchio del vaso 4 Mancata aspirazione Guarnizione del coperchio non in sede Svitare il coperchio e riposizionare la guarnizione nella sede del coperchio 5 Galleggiante bloccato Incro...

Страница 8: ...rocedere alle operazioni di pulizia come indicato nel capitolo pulizia ATTENZIONE Il lato del filtro di protezione contrassegnato dalla dicitura IN deve essere sempre connesso verso il vaso di aspiraz...

Страница 9: ...liance coming into contact with liquids Keep off the reach of children or not capable people without supervision Don t leave the appliance connected to the power supply socket when not in use Hold plu...

Страница 10: ...ment Fuse Alternate Current Battery Direct Current Hz Mains Frequency I ON O OFF DEHP Phthalates Suction catheter ACCESSORIES SUPPLIED DESCRIPTION COMPLETE ASPIRATION JAR 1000cc CONICAL FITTING TUBES...

Страница 11: ...gnetic field IEC EN 61000 4 8 3A m The power frequency magnetic field should be measured in the intended installation location to assure that it s sufficiently low Conducted Immunity IEC EN 61000 4 6...

Страница 12: ...in maximum vacuum position check that the vacuum indicators reaches 75 kPa 0 75 bar minimum internal battery Rotate the knob from right to left and check the aspiration regulating control The vacuum...

Страница 13: ...nnector 6 Connect the other end of the long silicon tube 5 to the probe plastic connector 8 then connect the suction probe to it Connect the power cord to the device then connect the plug to the elect...

Страница 14: ...sitionner l appareil sur des surfaces planes et stables Positionner l appareil de fa on viter d en occlure les prises d air sur la partie post rieure Ne pas utiliser l appareil en pr sence de m langes...

Страница 15: ...d aspiration ELIMINATION DES BATTERIES USAG S Directrive 2006 66 CE Applicable dans les pays de l Union Europ enne et aux autres pays europ ens disposant de syst me de collecte selective Ce symbole ap...

Страница 16: ...s des batteries Trous de tension br ves interruptions et variations de tension IEC EN 61000 4 11 5 UT for 0 5 cycle 40 UT for 05 cycle 70 UT for 25 cycle 5 UT for 5 sec L alimentation devrait tre cell...

Страница 17: ...autoclave o ils peuvent effectuer un cycle de st rilisation une temp rature de 120 C Le raccord conique qui vient fourni avec les tubes d aspiration peux tre utilis une temp rature de 121 C NE JAMAIS...

Страница 18: ...pire pas Joint du couvercle pas en place D visser le couvercle et remettre le joint en place 5 Le flotteur bloqu Pr sence des incrustations sur le flotteur D visser le couvercle et mettre la garniture...

Страница 19: ...ires et proc der aux op rations de nettoyage comme indiqu dans le chapitre nettoyage ATTENTION Le c te du filtre de protection marqu par la mention IN doit tre toujours branch vers le vase d aspiratio...

Страница 20: ...Ger t bei Nichtbenutzung nicht an der Netzsteckdose angeschlossen lassen Nicht am Netzkabel ziehen sondern den Stecker mit den Fingern festhalten um ihn aus der Steckdose zu ziehen Das Ger t gesch tz...

Страница 21: ...E Ger t der Isolierstoffklasse II CE Kennzeichnung gem Richtlinie 93 42 EWG und nachfolgende nderungen Achtung Im Handbuch nachlesen K hl und trocken lagern Lagertemperatur 40 C 70 C Ger t Typ B Siche...

Страница 22: ...ie nicht der Lebenserhaltung dienen Anmerkung UT ist der Wert der Einspeisungsspannung VORSCHRIFTEN F R ZUR CKSCHICKEN UND REPARIEREN In bereinstimmung mit den neuen EU Normen gibt der Hersteller hier...

Страница 23: ...it des Ger ts sind vor seiner Benutzung aber einige einfache Kontrollen auszuf hren Das Ger t auspacken und immer pr fen dass die Kunststoffteile und das Speisekabel unversehrt sind weil sie beim vorh...

Страница 24: ...Deckel des Beh lters abdrehen und den Beh lter zu 1 3 mit Wasser f llen um das Reinigen und den Vakuumaufbau w hrend des Betriebs zu beschleunigen Den Deckel dann auf den Beh lter aufschrauben Die Zu...

Страница 25: ...s El aparato solo se puede utilizar con filtros bacteriologicos Nunca sumergir el aparato en agua Posicionar el aparato sobre superficies planas y estables Posicionar el aparato de modo tal de evitar...

Страница 26: ...R UNI EN ISO 10079 1 ALTA ASPIRACION FLOJO BAJO Alimentac on 230V 50Hz o equipo internamente accionado bater a PB 12V 3Ah o 12Vdc 1 9A de la bater a de coche Potenzia Absorbida 50VA Fusible F 1 x 630m...

Страница 27: ...un ambiente comercial o de un hospital o ambientes a usted de las bater as Agujeros de tens on breves interrupciones y variaciones de tens on IEC EN 61000 4 11 5 UT para 0 5 ciclos 40 UT para 05 ciclo...

Страница 28: ...vac o o vertidos l quidos Los tubos de aspiraci n se pueden esterilizar en autoclave usando el programa de 120 C Los conectores conicos pueden ser esterilizados tambien en autoclave en el programa de...

Страница 29: ...n fuera del alojamiento Desenroscar el tap n y acomodar la junta en su alojamiento 5 Falta cierre del flotante Si el tap n ha sido lavado controlar que el flottante no se haya desenganchado parcialme...

Страница 30: ...cador IN o EN del filtro est en el lado que est enfrente de la cubierta del recipiente de acumulaci n y tambi n que est bien posicionado en el VAC O Una conexi n incorrecta causar la destrucci n immed...

Страница 31: ...z modyfikacji urz dzenia wynikaj cej z naprawy bez zezwolenia lub je eli kt rakolwiek z cz ci urz dzenia jest uszkodzona w wyniku wypadku lub niew a ciwego u ycia Wszelkie minimalne zmiany naprawy ur...

Страница 32: ...ch awaryjnych lub u pacjent w kt rych ryzyko infekcji nie jest znane filtr nale y zmieni po ka dym u yciu Filtry s dost pne w r nych wersjach wraz z pojemnikiem 2000cm Cewnik do odsysania cz jednorazo...

Страница 33: ...r 230 12V BR IEC EN 61000 3 2 IEC EN 61000 3 3 NEW ASKIR 230 12V BR NEW ASKIR 230 12V BR ESP IEC EN 61000 4 2 6kV 8kV 30 IEC EN 61000 4 4 2kV IEC EN 61000 4 5 1kV IEC EN 61000 4 11 5 UT 0 5 40 UT 05 7...

Страница 34: ...Upewni si czy zbiornik ustawiony jest w sterylizatorze dnem do g ry Umie ci cz ci w autoklawie i nastawi jeden cykl sterylizacji przy 121 C przy ci nieniu 1 bar Mechaniczna odporno zbiornika gwarantuj...

Страница 35: ...szkodzony Dreny zablokowane za amane lub od czone Zaw r odcinaj cy p ywakowy zablokowany uszkodzony silnik pompy Skr w prawo regulator i sprawdzi warto pr ni na skali Wymieni filtr Wymieni lub pod czy...

Страница 36: ...M 230V 50Hz 45 50 8 45 120 NIGDY NIE U YWA URZ DZENIA BEZ ZBIORNIK W I FILTRA OCHRONNEGO UPEWNIJ SI E DZIECI LUB OSOBY CHORE PSYCHICZNIE NIE U YWAJ URZ DZENIA BEZ NADZORU DOROS YCH ZAWSZE NALE Y UMIES...

Страница 37: ...rinde konumland r n Cihaz arka taraf ndaki hava giri leri engellenmeyecek bi imde konumland r n Hava oksijen veya nitrik oksit ile etkile ime girdi inde tutu abilir anestetik kar mlar bar nd ran ortam...

Страница 38: ...ESUARLAR A IKLAMA B T N ASP RASYON KAVANOZ 1000cc KON K BA LANTI ELEMANI T P SET 6 mm x 10 mm HAVALANDIRMA SONDASI CH20 ANT BAKTER YEL F LTRE Bu filtre s v lar n pn matik devre i ine giri ini nleyen h...

Страница 39: ...reklilik nitesinin kullan lmas tavsiye edilir Manyetik alan IEC EN 61000 4 8 3A m Yeterince d k oldu undan emin olmak ad na g frekans manyetik alan cihaz kurulumunun yap laca yerde l lmelidir letkenli...

Страница 40: ...ula t n do rulay n D meyi sa dan sola do ru evirin ve aspirasyon reg l syon kontrol ne dikkatedin Vakumg stergesi 25 kPa 0 25 bar d zeyine inecektir Cihazdan y ksek sesler gelmedi ine emin olun bunlar...

Страница 41: ...umunun daha h zl eri mesi ama lan r ve ard ndan kavanoz zerindeki kapa do ru bir ekilde tekrar s karak kapat n UYARI emin IN filtre marker taraf VACUUM i ine toplama kavanoz kapa ve monte d n k a k Bi...

Страница 42: ...Nutzzeit darf das Produkt NICHT zusammen mit dem Siedlungsabfall beseitigt werden Es kann zu den von den st dtischen Beh rden eingerichteten Sammelstellen oder zu den Fachh ndlern die einen R cknahme...

Страница 43: ...m the vector GARANT CA MI taraf ndan sunulan r nler sat n alma tarihinden itibaren 24 ay s reyle garanti kapsam nda yer al r Bu garanti do rultusunda fabrikam zda yahut yetkili Hizmet Merkezimizde Tek...

Страница 44: ...___________________________________________ Via Street Localit Place _______________________________________________________________________________________________________________________ Venduto A P...

Отзывы: