11

 Semi-Dry 

Blotters 

a.   Prepare 10 sheets of blotting paper cut to the same size of the gel and soak in 

0.1TXAE.  Cut a nylon membrane to the same size and wet in ddH

2

O for 5 

minutes followed by equilibrate in 0.1XTAE for 20-30 minutes. 

b.   Prepare one of the clear plastic Mylar sheets (provided in the kit) by cutting a 

hole in the sheet which is 1-2mm smaller than the gel dimension. 

c.   Insert the Mylar mask over the bottom electrode anode.  Place 5 pieces of 

saturated filter paper carefully centered over the cut-out in the Mylar mask.  If 
prepared correctly, the filter paper will overlap the mask cut-out slightly.  
Flatten the filter paper by rolling a glass rod over the surface.  Apply the 
equilibrated Nylon membrane to the stack of filter paper followed by the gel 
and 5 more pieces of saturated filter paper.  Again roll a glass rod over the 
filter paper stack to compress and remove trapped air bubbles.  The lid of the 
apparatus containing the cathode should be placed directly over the stack.  
While holding the two white nylon thumb screws, place the lid over the stack 
and secure while applying some downward pressure. 

 

4.   Electro-blotting transfer 

a.   Once the lid is in place, insert the power cords into the apparatus.  Plug the 

ends of the power cords into the Power Supply. 

b.   Set the Power Supply at 250mA and transfer for 30 minutes  at constant 

current. 

c.   After the run, turn off the power supply and disconnect the power cords from 

the apparatus.  The lid cannot be lifted unless the power cord is unplugged 
from the anode.  Carefully lift the lid and watch for filter paper sticking to the 
top electrode.  Remove the Nylon membrane and process according to 
protocol.  Rinse the inside of the apparatus and the mask with ddH

2

0.      

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 

Содержание EBU-4000

Страница 1: ...Semi Dry Blotting Systems INSTRUCTION MANUAL EBU 4000 EBU 6000...

Страница 2: ...ction 5 1 2 Specifications 5 1 3 Safety 5 Section 2 Description of parts 2 1 Unpacking EBU 4000 or 6000 6 2 2 Components Assembly 6 Section 3 Instructions for Use 3 1 Blotting Unit Preparation 7 11 Se...

Страница 3: ...e highest practical standards of materials workmanship and design C B S Scientific warrants that the product has been tested and will meet or exceed published specifications This warranty is valid onl...

Страница 4: ...endimiento insatisfactorio atribuible a circunstancias fuera del control de C B S Scientific C B S Scientific en ning n caso asumir responsabilidad por da os incidentales o subsecuentes incluyendo en...

Страница 5: ...t passes directly through the gel Use of the mask will increase transfer efficiency and reduce transfer times up to 50 1 2 Specifications Constructions Buffer chamber safety cover Acrylic Electrode pa...

Страница 6: ...nts Blotting chamber with stainless steel cathode and platinum coated anode Safety cover with power leads Mylar sheets for shielding uncovered portion of anode 2 2 Components Assembly SAFETY COVER POW...

Страница 7: ...is why nylon membranes are generally used uncharged nylon being preferable to positively charged nylon Nylon is the membrane of choice because it will bind DNA fragments as small as 50bp and can be ir...

Страница 8: ...g on thickness 3 6mm c Cut eight sheets of 3mm blotter paper Whatman the exact size of the gel Saturate all eight sheets in running buffer d Mylar Mask Prepare a mask by cutting a rectangular hole tha...

Страница 9: ...ng a rectangular hole that is slightly smaller 1 2mm than the gel dimensions b Prepare 8 pieces of Whatman 3mm chromatography paper the exact size of the gel Larger pieces of filter paper may promote...

Страница 10: ...At the conclusion of the run turn power supply off and disconnect leads from blotter Cover will not release unless power cord is removed Remove the cover slowly to catch any part of the gel sandwich...

Страница 11: ...lowed by the gel and 5 more pieces of saturated filter paper Again roll a glass rod over the filter paper stack to compress and remove trapped air bubbles The lid of the apparatus containing the catho...

Страница 12: ...sis 86 315 327 VCH Weinheim Germany 3 Dunn S D Effects of the modification of transfer buffer composition and the renaturation of proteins in gels on the recognition of proteins on Western blots by mo...

Страница 13: ...lpyrocarbonate DEPC treated water for extended periods at 37 C Electrostatic charge and heat affix the conductive platinum surface covering the titanium It can be scratched off or damaged by using sha...

Страница 14: ...t Item EBU 4000 Semi Dry Blotting System Includes power leads and safety interlock 20cm x 20cm EBU 6000 Semi Dry Blotting System Includes power leads and safety interlock 35cm x 45cm MM 4000 Mylar Mas...

Страница 15: ...NOTES...

Страница 16: ...d online ordering www cbsscientific com E mail address sales cbssci com Mailing address C B S Scientific Company P O Box 856 Del Mar CA 92014 Shipping address C B S Scientific Company 10805 Vista Sorr...

Отзывы: