C a l v a l u e s a r e
s e t t o d e f a u l t
O n l i n e M o d e
K e y b o a r d i n a c t i v e
Cause
The RAM and EEPROM contain
different calibration data, since
calibration data was not saved after
a change.
Solution
The calibration data is transferred
automatically from the EEPROM to
the RAM, and thus the old
calibration values are restored.
Cause
The instrument is in Online Mode,
and is therefore being controlled
from the PC. Only the <operate>
and <reference> buttons and the
foot switch are active.