5.3 Changing Calibration Values
The values for the calibration
standards are stored in the
instrument. In automatic calibrati-
on, this data is associated with the
respective standards.
In some cases it may be necessary
to adjust the stored data to a new
standard, for example if the
supplied standard has deteriorated
due to aging.
To input new standard values, select
the "Change Calibration
Value" function in the SET-UP menu,
and confirm with the
<operate> button.
5.4 Calibration with Taber
Abrasion Holder
s #ALIBRATETHEINSTRUMENTWITHOUT
the Taber Abrasion Holder.
s
Install the Taber Abrasion Holder in
front of the sphere entrance.
s
Calibrate the instrument once
more, but only dark- and bright
calibration. When you are asked to
calibrate Clarity, please press the
„set-up“ button.
s
Perform measurements without a
sample in Haze, Transmission and
Clarity. The result should be:
H = 0.00, T = 100 and C = 100.
If the Taber Abrasion Holder is remo-
ved, calibrate the instrument again.
Note:
When working with the Taber
Abrasion Holder, only measure
Haze or Transmission. Clarity measu-
rements are not possible
with the holder due to changed
optical conditons.
Содержание AT-4725
Страница 3: ......
Страница 45: ......
Страница 46: ...haze gard plus Best Nr AT 4725 Best Nr AT 4726 238 013 796 EDFI 0805...
Страница 47: ......
Страница 89: ......
Страница 90: ...haze gard plus N R f AT 4725 N R f AT 4726 238 013 796 EDFI 0805...
Страница 91: ......
Страница 133: ......
Страница 134: ...haze gard plus No di cat AT 4725 No di cat AT 4726 238 013 796 EDFI 0805...
Страница 135: ......
Страница 180: ...238 013 796 EDFI 0805...