GasAlertMicroClip
Charging the Rechargeable Battery
15
•
Force Calibration When Overdue:
When enabled, the
detector automatically enters calibration during startup for
overdue sensors. If the sensors are not calibrated
immediately, the detector will deactivate.
•
Cal Lock:
When enabled, the sensors can only be
calibrated using an IR device (IR Link with Fleet Manager
II or the MicroDock II base station).
•
Force Bump When Overdue:
When enabled, a bump test
is required if the sensor has exceeded its bump test
interval. If a successful bump test is not performed, the
detector will deactivate.
•
IntelliFlash:
When enabled, the green LED flashes to
provide continuous visual confirmation that the detector is
operating correctly.
IntelliFlash automatically deactivates during a low
battery alarm, a self-test fail, a calibration fail, a bump
test fail, or during an alarm event. To define how often
IntelliFlash occurs (
1
-
60
seconds), refer to Confidence
Beep and IntelliFlash Interval. Default setting is
1 second. Detector is shipped with IntelliFlash enabled.
•
Confidence Beep and IntelliFlash Interval:
Enter a
value (
1
-
60
seconds) to define how often IntelliFlash
occurs and the detector beeps. Intelliflash and/or
Confidence Beep must be enabled in order to define
Confidence Beep and IntelliFlash Interval.
•
Language:
Select the language to display on the LCD:
English
,
Français
(French),
Deutsch
(German),
Español
(Spanish), or
Português
(Portuguese).
Charging the Rechargeable Battery
a
Warning
Only the manufacturer can replace the rechargeable
battery. Failure to adhere to this caution can lead to fire
and/or explosion.
Charge only in a safe area that is free of hazardous gas
and within temperatures of 32°F to 113°F (0°C to 45°C).
The charging adapter is specific to your region. Use of the
charging adapter outside your region will damage the
charger and the detector.
Do not calibrate during or immediately after charging.
a
Avertissement
Seul le fabricant est habilité à remplacer la rechargeable
batterie. Le non-respect de cette consigne peut provoquer
un incendie et/ou une explosion.
Chargez le détecteur uniquement dans une zone sûre ne
renfermant pas de gaz dangereux et dont la température
ambiante reste comprise entre 0 à 32 à 113 °F (45 °C) To
charge the battery:
1.
Deactivate the detector. Insert the charging adapter
plug into an AC outlet.