GasAlertMicroClip
User Manual
No Gas Detected
When the detector senses a sufficient amount of
gas concentration (approximately 30 seconds), the
audible alarm beeps once,
flashes, and
K
remains lit while the detector completes the span
(approximately 2 minutes).
The LCD displays
FAILURE
if the detector does not detect
any gas within 2 minutes before proceeding with the
calibration procedure.
Note
Successful Span
The detector will not span a sensor if:
If the sensors have spanned successfully, the audible alarm
beeps before the calibration procedure continues.
x
You do not apply gas to the sensor.
x
The sensor fails to detect at least one-half of the
expected gas concentration in the first
30 seconds.
Unsuccessful Span
If the sensors fail the span, the LCD displays
FAILURE
before proceeding with the calibration procedure.
x
The gas concentration drops below one-half of
the expected gas level during the 2 minute span.
32