GasAlertMicroClip
User Manual
Note
The sensors are continuously tested while the
detector is activated.
Automatic Zero and O
2
Calibration (optional)
5.
The sensors (H
2
S, CO, and LEL) are automatically
zeroed and the O
2
sensor is automatically
calibrated (unless it is disabled in the detector’s
options). The detector beeps twice to signal a
successful zero and span.
If only the automatic oxygen calibration is enabled on
the detector, the following screen appears.
Note
If ambient air is set to be measured 20.8% vol., the
automatic oxygen calibration screen displays
20.8
instead of 20.9.
Calibration Due Date (optional)
6.
The LCD displays the number of days remaining
until the next calibration is due.
Note
If the
calibration interval
is set to 0, then the
calibration due date start-up feature is bypassed.
If any sensor is past its calibration due date, the
detector displays
CAL DUE todAY
as the detector
beeps, flashes, and vibrates.
12