5 | Installation
12 / 36
0870567093_NC0630C_ACV_A0002_IM_en
FS
Flow switch
MAN
Manometer
MV
Solenoid valve
NIP
Nitrogen panel
PRV
Pressure regulating valve
• Connect the barrier gas connection (BGC) to the gas supply.
Connection size:
– G1/4, ISO 228-1
Version with nitrogen panel:
• Electrically connect the solenoid valve (MV), see
Wiring Diagram Solenoid Valve
[
►
19]
.
• Electrically connect the flow switch (FS) of the flow meter, see
Wiring Diagram
Flow Switch [
►
21]
.
• Make sure that the gas complies with the following requirements:
Gas type
Dry nitrogen or air
Gas temperature
°C
0 ... 60
Maximum gas pressure
bar
13
Recommended pressure setting at the
pressure regulating valve (PRV)
bar
3
Filtration
µm
5
Recommended flow rate
SLM*
3.5 ... 5.5
Air quality (only for air)
Acc. to ISO
8573-1
Class 5.4.4.
* standard litre per minute
5.2.4 Dilution Gas System Connection (Optional)
PRV
MV
DGC
MAN
FME, FR
NIP
DGC
Dilution gas connection
FME
Flow meter
FR
Flow regulator
MAN
Manometer
MV
Solenoid valve
NIP
Nitrogen panel
PRV
Pressure regulating valve
• Connect the dilution gas connection (DGC) to the gas supply.
Connection size:
– G1/4, ISO 228-1
• Electrically connect the solenoid valve (MV), see
Wiring Diagram Solenoid Valve
[
►
19]
.
• Make sure that the gas complies with the following requirements: