11
Steam Sterilization
Minimum
Temperature
Minimum Exposure
Time
Dry Time
Pre Vacuum (Wrapped)
270° F (132° C)
3 Minutes
No Dry Time
Steam Gravity*
270° F (132° C)
3 Minutes
No Dry Time
Parameters for Steam Sterilization of BSPMAX™ II Battery Packs
Notes:
• These processes have been validated as being capable of cleaning and sterilizing BSPMAX™ II battery packs.
• The sterilizer manufacturer’s written instruction for cycle parameters, load configuration and AAMI guidelines
for steam sterilization should be followed.
Sterilization Recommendations (continued)
BSPMAX™ II battery packs have been validated for sterilization in a STERRAD
®
gas plasma system.
Sterilize per STERRAD
®
user instructions.
Note:
After minimum sterilization exposure time immediately remove battery pack. Allow to air dry and cool
for at least one hour before use.
*Unwrapped only
Other Sterilization Option
Содержание BSP MAX II PM-X00-710
Страница 1: ...PM X00 710 PM X00 715 IR 12 13 SURGICAL POWER ACCESSORIES...
Страница 2: ......
Страница 13: ...13 Return Goods Policy Contact your distributor regarding returned goods policy...
Страница 15: ......