Sterilization Recommendations
Steam sterilization has been found to be both safe
and effective for the sterilization of BSPMAX™ II Large
Bone Power System battery packs. The battery packs
are capable of withstanding the recommended
exposure times and temperatures of steam
sterilization.
Warning:
• The use of disinfecting solutions for an exterior
instrument wipe will not sterilize equipment.
Cautions:
• DO NOT sterilize battery packs with Ethylene Oxide
(ETO).
• DO NOT sterilize battery packs in cold sterilization
like CIDEX
®
.
• DO NOT “Peel Pack” battery packs for
sterilization. Sterilization in a sealed pouch traps
moisture which can cause damage.
• DO NOT sterilize battery pack while connected to
the handpiece.
• DO NOT sterilize a battery pack if damage (cracks
in the case) is apparent.
• DO NOT leave battery packs in a hot sterilizer for
more than the prescribed time as shortened
battery life will result.
• After sterilization, always allow battery packs to cool
for at least one hour before use.
Failure to comply may result in damage and a
shortened battery pack life.
• Charging of battery packs should be completed
prior to sterilization.