9
Cleaning Recommendations
Warnings:
• Clean, charge and sterilize battery packs before every
use.
Cautions:
• Follow universal precautions for protective apparel
when handling and cleaning contaminated battery
packs.
• Remove battery pack from contaminated
handpiece prior to cleaning and sterilization.
• DO NOT immerse a battery pack in any liquid.
Contact corrosion may occur resulting in decreased
performance of the battery pack and or handpiece.
• Do not clean battery packs in an automated
disinfecting washer or an ultrasonic cleaner.
• DO NOT use disinfectants with pH levels higher
than 10.5. Higher concentrations greater than
10.5 may cause cracks in the housing material.
• DO NOT clean the battery pack contacts with
abrasives.
• Dry battery packs before charging to avoid damage
to the battery packs and charger.
Cleaning Procedures:
1. Remove battery pack from the handpiece.
2. Inspect the battery pack case for cracks. Remove
damaged battery pack from service.
3. Wipe the battery pack case clean using a clean
lint-free soft cloth lightly dampened with a
hospital grade enzymatic disinfectant having a pH
range of 6.5 to 10.5.
4. Rinse the battery pack thoroughly under running
tap water.
5.
Dry the battery pack with a clean lint-free soft cloth.
6. Clean battery pack contacts using a cotton swab
lightly with isopropyl alcohol. DO NOT clean the
battery pack contacts with abrasives.
7.
Ensure battery pack and battery pack contacts are
completely dry before charging or using battery
pack.
8. Fully charge battery pack.
9. Sterilize
as
directed.
See
Sterilization
Recommendations section.