Cleaning Recommendations

Warnings:   

• Clean, charge and sterilize battery packs before every

use.

Cautions:

• Follow universal precautions for protective apparel

when handling and cleaning contaminated battery
packs.

• Remove  battery  pack  from  contaminated

handpiece prior to cleaning and sterilization.

• DO  NOT  immerse  a  battery  pack  in  any  liquid.

Contact corrosion may occur resulting in decreased
performance of the battery pack and or handpiece.   

• Do  not  clean  battery  packs  in  an  automated

disinfecting washer or an ultrasonic cleaner.

• DO NOT use disinfectants with pH levels higher

than  10.5.    Higher  concentrations  greater  than
10.5 may cause cracks in the housing material.

• DO  NOT  clean  the  battery  pack  contacts  with

abrasives.

• Dry battery packs before charging to avoid damage

to the battery packs and charger.

Cleaning Procedures:

1. Remove battery pack from the handpiece.
2. Inspect the battery pack case for cracks. Remove

damaged battery pack from service.

3. Wipe the battery pack case clean using a clean

lint-free  soft  cloth  lightly  dampened  with  a
hospital grade enzymatic disinfectant having a pH
range of 6.5 to 10.5.

4. Rinse the battery pack thoroughly under running

tap water.

5.

Dry the battery pack with a clean lint-free soft cloth.

6. Clean battery pack contacts using a cotton swab

lightly with isopropyl alcohol. DO NOT clean the
battery pack contacts with abrasives.

7.

Ensure battery pack and battery pack contacts are
completely dry before charging or using battery
pack.

8. Fully charge battery pack.
9. Sterilize 

as 

directed. 

See 

Sterilization

Recommendations section.

Содержание BSP MAX II PM-X00-710

Страница 1: ...PM X00 710 PM X00 715 IR 12 13 SURGICAL POWER ACCESSORIES...

Страница 2: ......

Страница 3: ...planation of Symbols Page 6 Features Page 6 Battery Pack Installation Removal Page 7 Troubleshooting Page 8 Cleaning Recommendations Page 9 Sterilization Recommendations Page 10 Specifications Page 12...

Страница 4: ...care cleaning sterilization and long term maintenance of BSPMAX II Large Bone Power System Introduction The BSPMAX II 9 6V Battery Pack PM X00 710 and 12V Battery Pack PM X00 715 are intended for use...

Страница 5: ...es Consult WHO and local regulations for further information Battery packs contain nickel cadmium which must be recycled or disposed of properly See product disposal instructions DO NOT store battery...

Страница 6: ...release the battery pack from the handpiece Battery Pack Latch Battery Pack Case Authorized representative in the European community Batch code Catalog number Caution Conforms with the essential requ...

Страница 7: ...battery pack is secure Rotate trigger to the RUN mode Test the handpiece and battery pack by depressing the trigger Rotate trigger to the SAFE mode when not in operation Battery Pack Removal Rotate t...

Страница 8: ...ttery pack End of life cycle for battery pack Replace battery pack Subjected to a dry cycle Replace battery pack 4 Bay Power Unit fuses are open blown Replace with appropriate rated fuses See4 Bay Pow...

Страница 9: ...battery pack contacts with abrasives Dry battery packs before charging to avoid damage to the battery packs and charger Cleaning Procedures 1 Remove battery pack from the handpiece 2 Inspect the batte...

Страница 10: ...sterilize battery packs in cold sterilization like CIDEX DO NOT Peel Pack battery packs for sterilization Sterilization in a sealed pouch traps moisture which can cause damage DO NOT sterilize batter...

Страница 11: ...MAX II battery packs The sterilizer manufacturer s written instruction for cycle parameters load configuration and AAMI guidelines for steam sterilization should be followed Sterilization Recommendati...

Страница 12: ...ully Charged 3 10 Minutes 3 15 Minutes Working Temperature 32 F 122 F 0 C 50 C 32 F 122 F 0 C 50 C Type Nickel Cadmium Ni Cd Nickel Cadmium Ni Cd Specifications Operating Ambient temperature 50 F to 7...

Страница 13: ...13 Return Goods Policy Contact your distributor regarding returned goods policy...

Страница 14: ...4 IN KEYLESS CHUCK PM X08 935 3 MM KEYLESS CHUCK PM X08 940 ZHS ZIMMER HUDSON STRYKER UNIVERSAL ATTACHMENTS REAMERS PM X08 911 1 4 IN JACOBS HI TORQUE W KEY EXTENDED LENGTH PM X08 950 HUDSON EXTENDED...

Страница 15: ......

Страница 16: ...stered trademark of Hudson Industries Inc Jacobs is a registered trademark of Jacobs Chuck Manufacturing Company Stryker is a registered trademark of Stryker Corporation Zimmer is a registered tradema...

Отзывы: