pH sensor - parameter
10
ph SenSor
-
pArAmeter
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
menu.
pH sensor MS01 | pH sensor < Parameter
Reference impedance
On
Diagnostics Maintenance
Filter response time
3s
Warning triggers
Error triggers
Hold
Restore factory values
Parameter
Function
pH sensor
Path to the menu or to the menu item
Fig. 5:
Parameter
menu of the function
pH sensor
The menu items are detailed in the following chapters:
•
Reference impedance
, refer to chpt.
•
Filter response time
, refer to chpt.
•
Warning triggers
,
Error triggers
, refer to chpt.
10.3, 10.4, 10.5, 10.6 and 10.10.
•
hold
, refer to chpt.
•
Restore factory values
, refer to chpt.
10.1
Deactivate or activate the measurement of the
impedance value of the reference electrode
The parameter
Reference impedance
makes it possible to activate or deactivate the measurement of the
impedance value. By default, the impedance is measured.
→
Make sure that the login user level is
Installer
. If necessary, change the user level. Refer to chpt.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Reference impedance
.
→
To deactivate the measurement of the impedance value, select
Off
.
→
To activate the measurement of the impedance value, select
On
.
→
Validate with
.
The impedance value is measured and displayed in the
Diagnostics
tab of the function
pH sensor
.
The impedance value can be monitored. Refer to chpt.