pH sensor - parameter
Do not configure and activate the low error limit.
3.
Configure and activate the high error limit of the leakage-current value.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Error triggers
.
→
Select
High ISFET leak. current
and enter a leakage-current value, above which an error event is gen-
erated. Validate with
.
→
Select
Activation flags
. Select
High ISFET leak. current
. Validate with
.
When the leakage-current value is more than the set limit, an error event is generated. Details about the
error event are given in chpt.
→
To enable the monitoring, i.e. to be informed when the leakage-current value is outside the normal range,
enable the diagnostics. See chpt.
10.7
Freeze the values that are transmitted on the fieldbus
The product makes it possible to temporarily freeze the values that are sent by the product on the fieldbus.
You can choose one of the following options:
•
If you want to manually freeze the values, refer to chpt.
•
If you want to automatically freeze the values, for example with a special event that is considered as a
trigger, refer to chpt.
. The special event is for example a cleaning operation of the product with the
Type MZ20 cleaning system.
10.7.1
Manual freeze
→
Make sure that the login user level is
Installer
. Refer to chpt.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
hold
1.
Choose the values to be sent on the fieldbus:
•
If you want to send the last-measured values, do the following procedure:
→
Select
Action
Last values
.
•
If you want to send user-specific values, do the following procedure:
→
Select
Action
User values
.
→
Select
pH
and enter a pH value. Validate with
.
→
Select
Temperature
and enter a temperature value. Validate with
.
English
Type MS01