78
Troubleshooting with messages
19.3.8
Message
Error: too high leakage current
Product-status symbol
Bit status of the PDO pH-state
bit 4 = 1
Possible cause
The leakage current of the measuring cell is above the set limit.
The message can only be displayed if the monitoring of the leakage
current has been configured and activated. Refer to chpt.
What to do?
→
Replace the measuring cell. Refer to chpt.
→
Check if the high error limit is consistent. Refer to chpt.
the high error limit is not consistent, send the product back to
Bürkert.
19.3.9
Message
Warning: too low pH
Product-status symbol
Bit status of the PDO pH-state
bit 7 = 1
Possible cause
The pH value of the water sample is is under the set limit.
The message can only be displayed if the monitoring of the pH
value has been configured and activated. Refer to chpt.
What to do?
→
Check the process.
19.3.10
Message
Warning: too high pH
Product-status symbol
Bit status of the PDO pH-state
bit 6 = 1
Possible cause
The pH value of the water sample is is above the set limit.
The message can only be displayed if the monitoring of the pH
value has been configured and activated. Refer to chpt.
What to do?
→
Check the process.
19.3.11
Message
Warning: too low temperature
Product-status symbol
Bit status of the PDO pH-state
bit 7 = 1
Possible cause
The temperature value of the water sample is under the set limit.
The message can only be displayed if the monitoring of the temper-
ature value has been configured and activated. Refer to chpt.
What to do?
→
Check the process.