24
pH sensor - parameter
10.2
Damp or not the variations of the measured
pH-values
The parameter
Filter response time
makes it possible to damp or not the variations of the measured pH-
values. If the variations of the measured values are not damped, then the raw measured values are sent on
the fieldbus.
By default, the parameter
Filter response time
is set to 3 s.
To change the value of the parameter
Filter response time
, do the following procedure:
→
Make sure that the login user level is
Installer
. If necessary, change the user level. Refer to chpt.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Filter response time
.
→
Enter a value between 0 s and 10000 s, depending on your installation. To not damp the variations of the
measured values, set the parameter to 0 s. The higher the time value is, the more the variations of the
measured values are damped.
→
Validate with
.
If the parameter is set to 0 s, then the variations of the measured values are not damped.
If the parameter is set to a value other than 0 s, then the variations of the measured values are damped.
10.3
Monitor the impedance value of the reference
electrode
The water pH is correctly measured if the reference electrode is in contact with the water sample. The
impedance value of the reference electrode shows whether the electrode is in contact or not with the water
sample:
•
If the impedance value is less than 160 k
W
, then the reference electrode is in contact with the water
sample;
•
If the impedance value is between 160 k
W
and 1 M
W
, then the reference electrode is not in contact with
the water sample.
You can be warned automatically when the impedance value of the reference electrode is not in contact
with the water sample. Do the following procedure:
1. Make sure that the login user level is
Installer
. Refer to chpt.
2.
Activate the measurement of the impedance value of the reference electrode. Refer to chpt.
Do not configure and activate the low warning limit.
3.
Configure and activate the high warning limit of the impedance value.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Parameter
.
→
Select
Warning triggers
.
→
Select
High ref. impedance
and enter the impedance value, above which a warning event is generated.
Validate with
.