background image

 

SMART

 

 
 

EBULLIOMETER

 

 
 
 
 
 

USER’S GUIDE 

 

 

Document revision: 11  

25.05.2022 

 

Firmware Version 01.11B 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание SMART EBULLIOMETER

Страница 1: ...SMART EBULLIOMETER USER S GUIDE Document revision 11 25 05 2022 Firmware Version 01 11B...

Страница 2: ...l 12 5 3 2 Manual reading of the boiling point and the percentage of alcohol 12 5 4 Settings menu 15 5 4 1 Login menu 15 5 4 2 Wine settings menu 16 5 4 2 1 Editing of Sugar content 17 5 4 2 2 Editing...

Страница 3: ...lock related errors 42 6 7 Description of error group 6 Power block and power management related errors 42 6 8 Description of error group 7 EEPROM related errors 43 6 9 Description of error group 8 Sy...

Страница 4: ...other components included in the sample 1 3 Working using classic ebulliometer When working with a classic ebulliometer the laboratory assistant must 1 To measure distilled water boiling temperature T...

Страница 5: ...Fig 1...

Страница 6: ...2 2 Measuring system Fig 2 2 3 Schematic of cooling system connections Fig 3...

Страница 7: ...2 4 Display and keyboard Fig 4 2 5 Rear panel Fig 5...

Страница 8: ...ure of sample measured 10 45 C 3 3 Power source Stationary 230 110 V AC 10 15 50 60 Hz Power consumption 200VA 3 4 Additional characteristics Overall dimension H x W x D 485 230 220 mm Weight 6 kg 4 I...

Страница 9: ...ar panel of the device Attention Power device from 3 prong grounded schuko plug only 5 Working with the device 5 1 Starting up After the connecting the power cable to the electrical network and turnin...

Страница 10: ...ystem should be flush with wine to be measured Close the discharge valve see Fig 2 and fill the measuring system to the top with wine to be measured about 100 ml Then open the discharge valve to evacu...

Страница 11: ...rs of the wine types see item 5 4 2 Wine type settings To select the wine type you will measure use buttons Arrow up Arrow down and then confirm with button OK Below is shown an example of the panel w...

Страница 12: ...ction for the selected wine type Alcohol content Calculated sample wine alcohol content Atmospheric pressure measured atmospheric pressure Sample boiling temperature measured sample wine boiling tempe...

Страница 13: ...ent is made on the basis of a manually read sample wine boiling temperature Wine type the selected wine type Note note for the selected wine type Sugar content sugar content for the selected wine type...

Страница 14: ...esults again press the PRINT button located on the front panel of the device or use the F6 key from the USB keyboard Notes 1 Please ignore the measurement result of the first measurement done after me...

Страница 15: ...ent in Water mode in the previous 24 hours a warning message will appear on the screen 5 4 Settings menu To start the Settings menu from the home screen select Settings icon using the buttons Left arr...

Страница 16: ...ermined 5 4 2 Wine settings menu From the Settings menu using the buttons Up arrow Down arrow select Wine settings menu item and confirm with OK button On the screen will appears a menu from which the...

Страница 17: ...ct which digit of the number to edit and use the buttons Down arrow Up arrow to set the desired value If a USB keyboard is connected to the device you can enter the desired value using the numbers 0 9...

Страница 18: ...ed value using the numbers 0 9 and To confirm the selected value press the button OK This way the value will be saved in the non volatile memory in the device If you do not want the changes to be save...

Страница 19: ...mples tested In case you are measuring dry wines only it is enough if at the end of the working day you run the device few times 2 3 in Cleaning mode using distilled water as a cleaning solution In ca...

Страница 20: ...hours to determine the volume of alcohol If you try to use the Measurement mode but you have not performed a measurement in the Water mode in the previous 24 hours a warning message will appear on the...

Страница 21: ...ased on multiple criteria After all criteria are met the measurement process is stopped the measured temperature is saved in the non volatile memory and the results are displayed as it is shown below...

Страница 22: ...temperature sensor calibration procedure includes several 2 preliminary distilled water measurements and several 3 distilled water measurements from which the new temperature sensor calibration value...

Страница 23: ...he Temperature sensor calibration option with Down arrow Up arrow and then confirm with the OK button The following message will appear on the display The following is an example panel of a started me...

Страница 24: ...lcohol solution used for calibration must have a nominal value in the range of 9 00 to 13 00 and a deviation from the nominal value no more than 0 02 The minimum amount of water alcohol solution requi...

Страница 25: ...e entered value must be in the range of 9 00 to 13 00 All measurements included in the calibration procedure including the preliminary one must be performed in strict compliance with the instructions...

Страница 26: ...will start with an information message of the type After completion of all measurements included in the calibration procedure the data from them are checked for repeatability and absolute accuracy If...

Страница 27: ...rrently selected language will be marked Use Down Arrow Up Arrow to select your preferred language and then confirm with the OK button Your selection will be stored in the non volatile memory of the d...

Страница 28: ...selected scale will be marked Use Down Arrow Up Arrow to select your prefered scale and then confirm with the OK button Your selection will be stored in the non volatile memory of the device and will...

Страница 29: ...allows you to select the type of external printer which the measurement results will be printed To enter in Printer menu from the Configuration menu see item 5 4 6 select the Printer menu option using...

Страница 30: ...or disable the beeps of the built in buzzer In case the buzzer is enabled the device emits characteristic beeps in the following cases After completing the start up procedure After completion of meas...

Страница 31: ...uted more than once a day In this case the measurements for one day will be saved in several files Therefore if you want to collect all the measurements for one day in one file you need to open and co...

Страница 32: ...inute encoding in the names of the data collection files M Minute 0 0 1 2 2 4 3 6 4 8 5 10 6 12 7 14 8 16 9 18 A 20 B 22 C 24 D 26 E 28 F 30 J 32 H 34 I 36 J 38 K 40 L 42 M 44 N 46 O 48 P 50 Q 52 R 54...

Страница 33: ...4 7 select the Data collection upload menu option using the Down arrow Up arrow buttons and then confirm with the OK A panel will appear on the display in which the user have enter the period of time...

Страница 34: ...lcohol Correction Alcohol correction for this wine type Wine Note Note for this wine type Sample Alcohol Content The calculated alcohol content of the sample Sample Boiling Temp C The measured boiling...

Страница 35: ...4 7 select the Error log upload menu option using the Down arrow Up arrow buttons and then confirm with the OK A panel will appear on the display in which the user have enter the period of time start...

Страница 36: ...up Error group Err Code Error code Example of opened file using text editor Error Log SN U0200001 FWV ET01 10B Date Time Err Group Err Code 21 09 2020 09 17 24 6 12 21 09 2020 09 48 08 6 4 More detail...

Страница 37: ...ate month and year for which he wants to upload the corresponding files In this panel use the Left arrow Right arrow buttons to select the field for the start or end date month and year Once you have...

Страница 38: ...iles Diagnostic files Note Considering the device keeps on SD card all these file types from the last 90 days so this operation may take time 5 4 7 5 Pressure calibration restore During the producing...

Страница 39: ...vice as well as in the factory database The same serial number is printed on a label on the back panel of the device In case of damage of the serial number stored in the device by some reason the user...

Страница 40: ...rsion The latest files with published firmware versions can be obtained from the distributors or manufacturer After receiving the firmware update file the user must copy this file to a USB flash pen d...

Страница 41: ...arrow buttons select the desired file and then confirm with the OK A panel with warning message will appear on the screen From this panel using Left arrow Right arrow select YES and confirm with OK bu...

Страница 42: ...er Firmware version Current firmware version Tpwr Current power block temperature Tpcb Current PCB temperature Bulteh 2000 Ltd device manufacturer www bulteh com manufacturer s website 5 4 10 System T...

Страница 43: ...Arrow buttons and then press the OK button Once the power block temperature drops below 45 0 C and the PCB temperature drops below 40 0 C the device automatically exits this mode and measurements are...

Страница 44: ...ccess denied due to prohibited access 9 The file directory object is invalid 10 The physical drive SD card USB flash memory is write protected 11 The logical drive number is invalid 12 The volume has...

Страница 45: ...wer block and power management related errors Code Description 1 Heater power value is out of range 2 Error in heater over current comparator 3 Heater over current registered from the comparator 4 Pow...

Страница 46: ...D is out of range 6 10 Description of error group 9 Atmospheric pressure sensor related errors Code Description 1 Wrong ID 2 Wrong structure 3 Device not found 4 Invalid length 5 Communication error 6...

Страница 47: ...range 11 Offset value error 12 Calibration Input temperature is too high 13 Calibration Unstable results 6 12 Description of error group 11 Cooling fan related errors Code Description 1 Speed measure...

Страница 48: ...t This warranty does not cover expenses incurred in connection with defects resulting from improper operation maintenance or abuse This Warranty Card has to be filled in at the time of purchase of the...