5.4.5. "Calibration" menu
To select the "Calibration" menu, from the "Settings" menu (see item 5.4.), Select the "Calibration"
option using <Down arrow> / <Up arrow> buttons and then confirm with the <OK> button.
To return to the
"Settings" Menu, press the <MODE> button. To return to the Home screen, press the <HOME>
button.
5.4.5.1. Temperature sensor calibration
Periodic calibration of the temperature sensor guarantees the correct operation of the SMART
EBULIOMETER and accurate measurement results.
The temperature sensor calibration procedure includes several (2) preliminary distilled water
measurements and several (3) distilled water measurements, from which the new temperature
sensor calibration value is calculated.
The temperature sensor calibration should be performed periodically (90 days or often, at the
discretion of the laboratory assistant).
Note: If the temperature sensor has not been calibrated in the previous 90 days, the following reminder
message will appear on the display when the "Measurement" or "Water" mode is started:
If you select "YES" with the <Left arrow >/ <Right arrow> buttons and confirm with the <OK > button, the
temperature sensor calibration procedure will start.