The diagnostic files have "EDG" file extension. They can not be open and they are intended to be
used for analysis from the producer.
Example of file name:
200908F4.EDG
where:
20 - year
09 - month
08 - date
F - hour ( encoded, in this example - 15)
4 - minute (encoded, in this example 8 )
5.4.7.4. Upload all files
This menu option allows you to upload(copy) all existing on SD card files to the external USB flash-pen
drive.
The files that will be upload are as follows:
- Data collection files
- Error log files
- Diagnostic files
Note: Considering the device keeps on SD card all these file types from the last 90 days, so this operation
may take time.
5.4.7.5. Pressure calibration restore
During the producing of each device the producer performs calibration of the build-in atmospheric pressure
sensor, using a high-precision barometer, keeping an archive copy of the calibration values in the factory
database.
In case of damage of the calibration values stored in the device by some reason, the user may request the
manufacturer to obtain a data file to restore the pressure calibration.
After receiving the file, the user must copy this file to a USB flash-pen drive in the CALDATA folder.