After flushing with 100 mL of water-alcohol solution, empty the measuring system and pour the water-
alcohol solution to the level line, confirm with the <OK> button to start the measurement.
After the completion of the preliminary measurement, several (three) similar measurements will be
performed, from which the calibration parameters will be calculated. Each subsequent measurement will start
with an information message of the type:
After completion of all measurements included in the calibration procedure, the data from them are checked
for repeatability and absolute accuracy.
If the results of all checks are within the acceptable limits, the calibration parameters are saved in the non-
volatile memory of the device and the following message is displayed on the screen:
5.4.6. "Configuration" menu
To enter in "Configuration" menu form the "Settings" menu ( see item 5.4.) select "Configuration" option,
using
the buttons <Up arrow> /<Down arrow> and confirm with <OK> button.