DeNOx
3 Transport and storage
The device should be only transported in the original case or in appropriate packing.
CAUTION
Transport
Transport the product in a way which is not damaging to health. Where necessary, use auxiliary means for transport and as-
sembly.
Avoid damage to the product. Treat the product with caution.
Ensure that the product is fastened with wall brackets which comply with DIN EN 61010-1.
If the device is not used for some time, protect it against heat and humidity. Store the device in a roofed, dry, and dust free
room. Temperature should be between -20 °C and 60 °C (-4 °F and 140 °F).
4 Installation and connection
4.1 Installation site requirements
Sample gas probes are intended for flange mounting.
• Installation site and installation position are determined based on requirements specific to the application.
• If necessary, the connection piece should be slightly tilted toward the centre of the channel.
• Installation site must be protected from the elements and should provide protection against sun and rain.
• In addition, adequate and safe access for installation and future maintenance work should be provided. Particularly follow
the uninstalled size of the probe tube!
• Never exceed the approved ambient temperature range (T
amb
) of -20 °C to +35 °C.
• The dew point must always be at least 5 K above the ambient temperature.
• The device must be protected from shock and impact.
If the probe is transported to the installation site in pieces, it will first need to be assembled.
4.2 Installing the sampling tube (optional)
The sampling tube, if necessary with the fitting extension, must be screwed in. The probe is then attached to the mating flange
using the included seal, screws and nuts.
4.3 Installing the outlet filter
NOTICE
The outlet filter and the O-ring for the handle are included loose and must be inserted prior to first startup.
Operating without outlet filter prohibited!
O-ring handle
4.4 Insulation
On heated probes completely insulate any exposed flange areas and, if applicable, the connection piece to absolutely prevent
thermal bridges. The insulating material must meet the application requirements and be weatherproof.