16 

16 

  

 

 

A/C Power Supply (continuous sampling) 

The Standard Battery  Charger for BUCK pumps, will supply 
enough power to operate the Libra Plus Pump continuously while 
sampling. Simply plug the charger into the battery pack into 120 
VAC outlet and turn the pump on to sample. 

Interchangeable Batteries 

The BUCK-Libra Series incorporates a unique system of inter-
changeable batteries to provide additional flexibility in sampling. 
A Triple Pack is available for extended run times. All run times 
are tripled from those stated for a single Standard Pack. 

 

Battery Pack 

The  BUCK-Libra Series Standard Battery Pack consists of four 
nickel metal hydride batteries. The batteries are rated at 2150mA 
hour capacity. The QuickOne™ and QuickFive™ Battery Chargers 
connect to the back of each battery case; by lifting the protective 
rubber cover, the charger lead can be easily inserted. A Triple  
Battery Pack may be substituted for the standard battery pack to 
provide three times as much battery capacity for the Libra Plus. 
When a Battery Pack or pump with Battery Pack is connected  to a 
BUCK QuickOne™ or QuickFive™ Charger,  
a yellow LED
(CHARGING) will indicate the charging cycle has begun. The 
charge cycle takes approximately one hour for the single packs 
and 3 hours for the triple packs. When the charging cycle is    
completed, a green LED will light (READY) indicating end of cycle. 
Individual battery packs may be charged independently of the 
pump if required. The pump may be left connected to the charger 
in “trickle charge” indefinitely. Battery pack cases are connected 
to the pump case bottom with four self-retaining screws. 

Caution:  Never charge batteries in hazardous areas. 

  

Section 4 

Содержание BUCK-Libra Plus LP-5

Страница 1: ...00 330 2825 407 851 8602 Fax 407 851 8910 apbuck apbuck com www apbuck com MAN LIBRA PLUS Dated February 8 2006 BUCK Libra Plus MODEL LP 5 7101 Presidents Drive Suite 110 Orlando FL 32809 Tel 407 851...

Страница 2: ...2 2 23 23...

Страница 3: ...Specification 7 8 4 Operational and Menu Displays 9 11 5 Filter Cassette Sampling 12 6 Bubble and Impingers Sampling 13 7 Low Flow Tube Sampling 14 8 Battery Chargers and A C Power Supply 15 9 Parts a...

Страница 4: ...on line at www apbuck com on the service and support page Please obtain an RMA number prior to returning any product Call us at 407 851 8602 and have the serial number and model of the unit available...

Страница 5: ...w adjust Menu 4 Press and hold the SET key and use arrows to adjust flow This clears all previous data 5 Releasing the SET key will store the flow rate 6 Press the ON OFF key to return to Main Display...

Страница 6: ...volume and flow rate Accuracy 5 of display reading or pump Flow Faults Count down timer up to 40 hours turns off pump Key pad lock system One year warranty 19 19 SERVICE INFORMATION For all work not...

Страница 7: ...or alleged to have risen from any delay in delivery malfunction or failure of the equip ment The Seller s liability for any other loss or damage arising out of or connected with the manufacture sale...

Страница 8: ...the triple pack 4 BUCK One Hour Rechargeable Standard Battery Pack APB 109023 for the Libra Plus Series of Pump These self contained packs may be charged independently from the pump and used as addit...

Страница 9: ...Chargers connect to the back of each battery case by lifting the protective rubber cover the charger lead can be easily inserted A Triple Battery Pack may be substituted for the standard battery pack...

Страница 10: ...ow and finished or ready Green The QuickOne and QuickFive Chargers will charge a completely discharged pump battery in one 1 hour Three 3 hours for the optional Triple Pack with a fully discharged pac...

Страница 11: ...gh the adsorbent tube as shown above Use a screw driver to adjust the flow to the desired rate 3 Once the sample flow rate has been set press the SET and Arrow key to match the measured flow on the Ca...

Страница 12: ...gins to flow at preset flow rate from the steps above 4 Measure flow if it matches 5 of setting press the ON OFF key and resume sampling Press and hold the SET key to adjust the Factor for the pump sp...

Отзывы: