Section 1
Buck-Libra Plus Series Quick Start Guide
Note: Connect a representative sampling filter cassette using 1/4
inch tubing to the pump prior to setting Flow Rate for optimal per-
formance
.
Main Display
Setting the Flow rate
1. Connect sampling filter with a hose to the pump.
2. Turn on the pump by pressing the ON key.
3. Next press the down arrow to the Flow adjust Menu.
4. Press and hold the SET key and
use arrows to adjust flow.
(This clears all previous data).
5. Releasing the SET key will store the flow rate.
6. Press the ON/OFF key to return to Main Display, then
press the up arrow (RUN) to begin sampling.
Pump calibration Adjustment if needed at point of Flow Set
1. To measure the flow rate have a Calibrator ready and con-
nected to the filter as shown below.
2. From the Main Display down arrow to the Calibration
Mode.
3. Press the SET key and release, pump
begins to flow at preset flow rate from
the steps above.
4. Measure flow, if it m/- 5% of setting, press the ON/
OFF key and resume sampling. Press and hold the SET
key to adjust the Factor for the pump speed to match de-
sired flow rate.