6 Operation
22
F-100 / F-105
Operation Manual, Version C
6 .5
Working with BUCHI vacuum controller
The Chiller model F-105 is equipped with a communication socket. Here, a BUCHI Rotavapor setup
or other appropriate laboratory instruments equipped with a BUCHI vacuum controller (model V-850
or higher with firmware version 3.0 or higher) can be attached.
To establish communication switch on all connected devices. After bootup time the UP and DOWN
buttons of the Chillers operation panel are blocked – the Chiller can be controlled via the Vacuum
Controller buttons.
*click*
To Rotavapor /
Vacuum pump
≤
5 m
NOTE
Do not exceed max. recommended cable length (5 m). For more information about operation see
respective manual of the Vacuum Controller in use.
Communication F-105
Vacuum Controller
V-850 or V-855
Interface I-100
Interface I-300 with
Legacy box
Start or stop chiller via controller or interface
Yes
Yes
Yes
Display set point and current temperature
Yes
No
Yes
Adjust set point via controller or interface
Yes
No
Yes